PNA Series Microwave Network Analyzers
Tests and Adjustments
E8361A
Non-ANSI/NCSL Z540–1–1994 Verification
Non-ANSI/NCSL Z540–1–1994 Verification
For non-ANSI/NCSL Z540-1-1994, perform the preliminary checks and either the system
verification or performance tests while stopping to troubleshoot. Refer to
Figure 3-2
for test
flow. Troubleshoot and repair the first problem encountered without continuing to other
tests. After you troubleshoot, repair, and adjust, repeat the last failed portion and generate
a new set of data. Print out the system verification results or complete the performance
test record as the analyzer passes each test.
Figure 3-2
Non–ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart
Service Guide E8361-90001
3- 9