Checking The Serial (Rs-232), Parallel (1284-C), Or Vga Port - Agilent Technologies PNA Series Service Manual

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

Troubleshooting
Rear Panel Troubleshooting

Checking the SERIAL (RS-232), PARALLEL (1284-C), or VGA Port

To verify the proper operation of the SERIAL, PARALLEL, or VGA port:
• Connect a known good serial, parallel, or VGA peripheral device.
• Wait 15 seconds for the analyzer to verify the device connection, and then check the
operation of the peripheral device.
• If the peripheral device performs correctly, the port is functioning properly.
• If the peripheral device does not function properly, the A15 CPU board is faulty. Refer to
"Removing and Replacing the A15 CPU Board" on page
4-22
PNA Series Microwave Network Analyzers
7-28.
Service Guide E8361-90001