PNA Series Microwave Network Analyzers
E8361A
If the System Fails the Verification Test
IMPORTANT
Inspect all connections. Do not remove the cable from the analyzer test port.
This will invalidate the calibration that you performed earlier.
1. Disconnect and clean the device that failed the verification test.
2. Reconnect the device making sure that all connections are torqued to the proper
specifications.
3. Measure the device again.
4. If the analyzer still fails the test, check the measurement calibration by viewing the
error terms as described in
5. Refer to
Figure 3-12
Figure 3-12 System Verification Failure Flowchart
Service Guide E8361-90001
"Accessing Error Terms" on page
for additional troubleshooting steps.
Tests and Adjustments
System Verification
A-6.