Performance Tests (Agilent N2721A Software Package); Source Power Accuracy Test - Agilent Technologies PNA Series Service Manual

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

Tests and Adjustments

Performance Tests (Agilent N2721A Software Package)

Performance Tests (Agilent N2721A Software Package
The Agilent N2721A software package verifies the electrical performance of your E8364B
PNA series microwave network analyzer. The software automatically configures your
analyzer to execute the performance tests. The N2721A software package is not included
with the analyzer; it must be ordered separately. The model numbers of the equipment
used are specified under
There are nine tests in the software package:

Source Power Accuracy Test

Source Power Linearity Test
Frequency Accuracy Test
Trace Noise Test
System Compression Test
Noise Floor Test
Calibration Coefficient Test
System Crosstalk Test
Dynamic Accuracy Test
Source Power Accuracy Test
Function of the Test: To confirm the accuracy of the source output power of your network
analyzer over its full frequency range.
Specification Tested: Test Port Output–Power Level Accuracy
Equipment Used: A power meter, power sensors, and adapters.
Description of the Test:
1. The analyzer's output power level is set to 0 dBm.
2. A power sensor is connected to Port 1.
3. The output power is measured at hundreds of CW frequencies across the analyzer's
frequency range and the values compared to the setting of 0 dBm.
If the Analyzer Fails this Test:
• Perform the
"Source Calibration Adjustment" on page 3-42
• If the analyzer still fails this test, replace the source assembly and repeat this test.
Refer to
"Removing and Replacing the A12 Source 20" on page
1. Available in June 2003.
3-28
"Required Service Test Equipment" on page
PNA Series Microwave Network Analyzers
2-6.
and repeat this test.
7-22.
Service Guide E8361-90001
E8361A
1
)