Tests and Adjustments
System Verification
4. In the Calibration Kit box, select the calibration kit that is being used by clicking on it.
The corresponding verification kit to use is selected for you and displayed in the
Verification Kit box. Refer to
5. Under Printer Output, click one of the following options. Refer to
• None: No printout of results.
• Tabular Data: Prints the verification data in tabular form which includes measured
data and uncertainty limits. For an example, refer to
• Measurement Plots: Prints the verification data in graphical form. The graphical
form includes the measured data trace, factory supplied data trace, and uncertainty
limits. For an example, refer to
• Both: Prints the verification data in both formats.
For printed output, it is assumed that the printer has been tested and the
NOTE
Windows 2000 driver is installed for the printer that is being used. The
system verification test prints to the printer that has been designated as the
default printer. (On the Windows Desktop display, click on My Computer,
Control Panel, and then Printers to verify the printer setup.)
6. Click Run.
7. Follow the instructions on the analyzer for performing a full 2-port calibration.
8. After completion of the full 2-port calibration, follow the instructions on the analyzer for
performing the system verification.
For the system verification, insert the devices as shown in
Figure 3-11 System Verification Device Connections
3-24
Figure
3-10.
Figure 3-14 on page
PNA Series Microwave Network Analyzers
Figure
3-10.
Figure 3-13 on page
3-27.
Figure
3-11.
Service Guide E8361-90001
E8361A
3-26.