Transmission Tracking (Etf And Etr - Agilent Technologies PNA Series Service Manual

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

Error Terms
Error Term Data
Transmission Tracking (E
Transmission tracking is the difference between the frequency response of the reference
path (including the R input) and the frequency response of the transmission test path
(including the A or B input) while measuring transmission. The response of the test port
cables is included. These terms are characterized by measuring the transmission (S
of the "thru" configuration during the measurement calibration.
All transmission measurements are affected by transmission tracking errors.
Typical Cause of Failure
The test port cable is the most common cause of transmission tracking specification failure.
If the test port cable performance has been independently verified, suspect the following:
• If both E
and E
TF
TR
— suspect the analyzer switch splitter
• If one of the track term specification fails
— suspect the coupler, step attenuator, or the test receiver of the failed port
To troubleshoot, refer to
Signal Separation Group" on page
Figure A-7 Typical E
A-14
and E
TF
fail
"Checking the Receiver Group" on page 4-52
4-49.
/E
TF
TR
PNA Series Microwave Network Analyzers
)
TR
Service Guide E8361-90001
E8361A
, S
)
21
12
or to
"Checking the

Advertisement

Table of Contents
loading

Table of Contents