Agilent Technologies PNA Series Service Manual page 55

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8361A
Figure 3-14
shows an example of typical verification results with Measurement Plots
selected in the Printer Output area of the System Verification windows. The printed
graphical results show the following:
• the name of the device measured
• the serial number of the device
• the parameters measured
• Results of the measurements. Labeled as A in
• Data measured at the factory from the verification kit. Labeled as B in
• Upper and lower limit points as defined by the total system uncertainty system.
Labeled as C in
Figure
Figure 3-14 Example of Printed Graphical Verification Results
Service Guide E8361-90001
3-14.
Figure
3-14.
Tests and Adjustments
System Verification
Figure
3-14.
3- 27