Using Error Terms As A Diagnostic Tool; Preventive Maintenance; Troubleshooting - Agilent Technologies PNA Series Service Manual

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8361A

Using Error Terms as a Diagnostic Tool

By examining error terms, you can monitor system performance for preventive
maintenance and troubleshooting purposes.
The magnitude and shape of the error terms are affected by:
• calibration kit devices
• cables
• adapters and accessories
• the assemblies from the signal separation group of the analyzer
Calibration kit devices, cables, and adapters and accessories are the most common cause of
error term anomalies. Make sure of the following:
• Connectors must be clean, gaged, and within specification.
• Use proper connection technique during measurement and calibration. For information
on connection technique and on cleaning and gaging connectors, refer to
Principles of Connector Care" on page
guide.

Preventive Maintenance

If you print or plot the error terms at set intervals (weekly, monthly and so forth), you can
compare current error terms to these records. A stable system should generate repeatable
error terms over long intervals, (for example, six months). Look for the following:
• A long-term trend often reflects drift, connector and cable wear, or gradual degradation,
indicating the need for further investigation and preventive maintenance. Yet, the
system may still conform to specifications. The cure is often as simple as cleaning and
gaging connectors and cables.
• A sudden shift in error terms may indicate the need for troubleshooting.

Troubleshooting

You can use the error terms as a tool to isolate faulty assemblies in the signal separation
group of your analyzer. You can compare the current values to preventive maintenance
records or to the typical values listed in
start in
"Error Term Data" on page
To find assemblies related to error term failures, refer to plot descriptions in
Data" on page
A-8. Each plot description lists common assemblies related to each error
term. Identify the assembly and refer to
NOTE
Always suspect calibration devices, cables, or improper connector
maintenance as the primary cause of an error term anomaly.
Service Guide E8361-90001
3-5or to the calibration kit's user's and service
Table A-1 on page
A-8.
Chapter 4, "Troubleshooting."
Using Error Terms as a Diagnostic Tool
"Review the
A-9, or data trace plots that
"Error Term
Error Terms
A- 3

Advertisement

Table of Contents
loading

Table of Contents