GE moisture.IQ User Manual page 115

Moisture and oxygen analyzers
Hide thumbs Also See for moisture.IQ:
Table of Contents

Advertisement

Background Gas Compatibility
STAB-EL® cell: All gas compositions including those containing "acid" gases such as CO
SO
, etc.
2
Hazardous Area Classification
BASO1ATEX1O98X
II 1G Ex ia IIC T5 Ga
T
= –4°F to +122°F (–20°C to +50°C)
amb
US/CAN Class I, Division 1, Groups A, B, C, D, T4
European Compliance
Complies with EMC Directive 2004/108/EC when connected to a moisture.IQ, Moisture Image
®
Moisture Image
Series 2 or Moisture Monitor™ Series 3 analyzer
Sample Requirements
Inlet Pressure
0.2 psig to 1.0 psig (0.013 barg to 0.06 barg) (standard range)
Flow Rate
0.5 to 1.5 SCFH
Moisture
No limits (avoid condensation)
Oil/Solvent Mist
Less than 0.5 mg/ft
Greater than 0.5 mg/ft
Solid Particles
Less than 2.0 mg/ft
Greater than 2.0 mg/ft
moisture.IQ User's Guide
3
(standard range)
3
(use filter)
3
(standard range)
3
(use filter)
Chapter 9. Specifications
, H
S, C
, NOx,
2
2
l2
®
Series 1,
105