Checking The Electrolyte Level; Replenishing The Electrolyte - GE moisture.IQ User Manual

Moisture and oxygen analyzers
Hide thumbs Also See for moisture.IQ:
Table of Contents

Advertisement

Chapter 7. Maintenance

7.1.1 Checking the Electrolyte Level

Using the level indicator window on the oxygen cell, check the electrolyte level. The electrolyte should cover about
60% of the window. The electrolyte level should be between the
When the Delta F Oxygen Cell connections are made to the moisture.IQ, the analyzer can detect the electrolyte level
and indicate when it is low. When the electrolyte level is low, all oxygen readings are highlighted in amber and an
"Oxygen Fluid Low" error message is displayed beneath the measurements.

7.1.2 Replenishing the Electrolyte

Once the oxygen cell receives the initial charge of electrolyte, you should monitor the level regularly.
fluid level drop below the
WARNING! The electrolyte contains a strong caustic component and can be harmful if it comes in
contact with the skin or the eyes. Follow proper procedures for handling the caustic (potassium
hydroxide) solution. Consult your company safety personnel.
To raise the fluid level in the reservoir, add
distilled water, making sure you do not overfill the reservoir. The electrolyte mixture should cover approximately 60%
of the level indicator window.
86
Figure 75: Electrolyte Level for the Delta F Oxygen Cell
MIN
level mark on the window.
DISTILLED WATER
Min
Max
and
lines, as shown in Figure 75 below.
Level
Indicator
slowly in small amounts. Check the level as you add the
DO NOT
let the
moisture.IQ User's Guide