GE moisture.IQ User Manual page 145

Moisture and oxygen analyzers
Hide thumbs Also See for moisture.IQ:
Table of Contents

Advertisement

Function
Parameter
5100 Single
Measurement
Read
Channel 1
moisture.IQ User's Guide
Table 13: Modbus Register Map (cont.)
Sub Parameter
ID
ModBus
Address
Block Write
0
Unit Codes
Hygro - Dew
2
Point °C
Hygro - Dew
4
Point °F
Hygro - Equiv.
6
Dew Point °C
Hygro - Equiv.
8
Dew Point °F
Hygro - PPMv
10
Hygro - PPBv
12
Hygro - PPMw
14
Hygro - RH %
16
Hygro -
18
Lbs/MMSCF
(ideal gas)
Hygro -
20
Lbs/MMSCF
(natural gas)
Hygro - PPMv
22
(natural gas)
Hygro - g/m^3
24
Hygro -
26
mg/m^3
Hygro - Vapor
28
Pressure (kPa)
Hygro - Vapor
30
Pressure (mm
Hg)
Hygro - MH / FH
32
Type
Access Modbus
5000 32-bit Integer
5102 32-bit float
RO
5104 32-bit float
RO
5106 32-bit float
RO
5108 32-bit float
RO
5110 32-bit float
RO
5112 32-bit float
RO
5114 32-bit float
RO
5116 32-bit float
RO
5118 32-bit float
RO
5120 32-bit float
RO
5122 32-bit float
RO
5124 32-bit float
RO
5126 32-bit float
RO
5128 32-bit float
RO
5130 32-bit float
RO
5132 32-bit float
RO
Appendix D. Modbus Register Map
Qty. of
Notes
Request
Registers
Write
2-32
Set Unit Codes for
Multiple
Measurements 0-15
(0x10)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
Read Input
2
(0x04)
135