GE moisture.IQ User Manual page 139

Moisture and oxygen analyzers
Hide thumbs Also See for moisture.IQ:
Table of Contents

Advertisement

Function
Parameter
Channel 6 /
Output B
3000 All Alarm Status
Channel 1 /
Alarm A
moisture.IQ User's Guide
Table 13: Modbus Register Map (cont.)
Sub Parameter
ID
ModBus
Address
Units
604
Type
606
Zero
608
Span
610
Output (percent
620
of scale)
State
622
Units
624
Type
626
Zero
628
Span
630
0
Status
100
State
102
Units
104
Type
106
Lower Setpoint
108
Type
Access Modbus
2604 32-bit Integer
RW
2606 32-bit Integer
RW
2608 32-bit float
RW
2610 32-bit float
RW
2632 32-bit float
RW
2622 32-bit Integer
RW
2624 32-bit Integer
RW
2626 32-bit Integer
RW
2628 32-bit float
RW
2630 32-bit float
RW
3012 32-bit Integer
RO
3100 32-bit Integer
RO
3102 32-bit Integer
RO
3104 32-bit Integer
RW
3106 32-bit Integer
RW
3108 32-bit float
RW
Appendix D. Modbus Register Map
Qty. of
Notes
Request
Registers
Read Holding
2
Measurement/Unit
(0x03)
Code 0 = off,
others See tab
Read Holding
2
1 = 4-20 mA,
(0x03)
2 = 0-20 mA,
3 = 0-2 volts
Read Holding
2
(0x03)
Read Holding
2
(0x03)
Read Holding
2
When TEST MODE
(0x03)
enabled, force output
to value
Read Holding
2
2 = TEST MODE,
(0x03)
All else = run mode
Read Holding
2
Measurement/Unit
(0x03)
Code 0 = off,
others see tab
Read Holding
2
1 = 4-20 mA,
(0x03)
2 = 0-20 mA,
3 = 0-2 volts
Read Holding
2
(0x03)
Read Holding
2
(0x03)
Read Input
2
Bit wise Alarm Status.
(0x04)
1 = Energized
0 = De-energized
Read Input
2
1 = Energized
(0x04)
0 = De-energized
Read Holding
2
0 = Disabled,
(0x04)
1 = Enabled
Read Holding
2
Measurement/Unit
(0x03)
Code 0 = off,
others see tab
Read Holding
2
1 = Setpoint,
(0x03)
2 = In Band,
3 = Out Band
Read Holding
2
(0x03), Write
Multiple
(0x10)
129