Chapter 4. Using the Outputs, Alarms and Logger Menus
4.1.2 Testing the Selected Output
Make sure the output device is connected to the moisture.IQ as described in Chapter 1 and complete the following steps
to test the output:
Channel (1-6)
1. Tap the
Figure 44 below.
2. Tap the text box and enter the desired Percent of Scale test value (0-120). Then, tap the
the test value to the output.
3. After about 5 seconds, this reading should be as shown in Table 7 below.
4. Repeat steps 1-3 for as many different test values as desired. When you have finished testing the output, tap the
Exit
button to return to the Output Selection screen.
48
Output
A
B
and
(
or
) to be tested, and then tap the
Figure 44: Output Test Screen
Table 7: Expected Multimeter Test Readings
Output Range
0 to 20 mA
4 to 20 mA
0 to 2V
Test
button to open a screen similar to
Multimeter Reading
20 x Test %/100 mA
4 + 16 x Test %/100 mA
2 x Test %/100 V
Check
button to send
moisture.IQ User's Guide