GE moisture.IQ User Manual page 30

Moisture and oxygen analyzers
Hide thumbs Also See for moisture.IQ:
Table of Contents

Advertisement

Chapter 1. Installation and Wiring
1.9.3b
Moisture Image Series Probes (cont.)
If the Moisture Image Series probe requires assembly, complete that task before proceeding. Then, connect the probe
cable to the terminal block labeled
You can connect the Moisture Image Series Probe to any channel. However, if you are also using other sensors such as
M-Series probes, be sure to connect the Moisture Image Series probe to an unassigned channel.
IMPORTANT:
Check the Calibration Data Sheets of all installed sensors to determine which channels already have
probes assigned to them.
Note: Additional MIS probes may be connected to
side of the back panel.
After you make the back panel connections, connect the other end of the probe cable to the Moisture Image Series
probe per the wiring diagram in Figure 17. Trim all other flying leads from the cable flush with the jacket.
After you complete the Moisture Image Series probe connections, you must activate the probe on the installed channel
as described in "The Probe Configuration Screen" on page 58.
20
MIS
on the back panel of the electronics unit (see Figure 16 below).
Module A - CH1
Module A - CH2
Module A - CH3
Figure 16: MIS Probe Connections
Figure 17: MIS Probe Cable Wiring Diagram
MIS Probe Connections
CH4
CH5
CH6
,
and
in the
MIS
1
2
MISP 1
or MISP 2
Module B
connector group on the right
moisture.IQ User's Guide