6 Operation
6.6 Programs
Explanation
Programming step S1
40
6.6.4

Starting the program

➢ Press the START/STOP key to start the process.
 The LCD displays shows the stand-by screen. In the top line, the
status READY changes to RUN. Blinking or illumination of the stage
LCD indicates that a process is running. Passed program stages are
no longer lit when they are completed.
You can pause the current program by pressing the START/STOP button
again.
6.6.5

Starting a program with the Casting-Time function

1. Select the desired program with the Load Program [ → 39] key.
2. Press the Casting-Time key.
 The LCD display shows the following screen:
1. Press the S1 key to set the desired casting time of the program.
2. Enter the day of the week with the numeric keys (MO=1, TUE =2,
WED=3 ..., SUN=7)
3. Enter the desired casting time (hh:mm) with the numeric keys.
4. Press the START/STOP key.
 The LCD display now indicates the program start in line "S1." The
casting-time function is now activated; there is no need to press any
further key for confirmation.
To quit the program press the Casting-Time key again.
6.6.6

Freely programmable programs

This section explains the freely programmable settings.

Entering program values

6.6.6.1
1. Press the S1 key (S1 key for step S1).
 The cursor will blink in the "Heating and cooling down speed"
column.
2. Enter the heating velocity with the numeric keys. This can range from
1 ℃ /min (33.8 °F/min) to 30 ℃ /min (86 °F/min). If the heating
velocity is 30 ℃/min (86 °F/min), the furnace will heat up at maximum
power.
3. The cursor now switches to the "holding temperature" column. Enter
the four digits of the holding temperature in °C/°F (e.g. 1540/2804).
NOTICE
Entries with less than four digits
If the entry only consists of three digits (or less), the cursor must be
moved to the "Holding time" column by pressing the corresponding S
key.
4. Enter the holding time in minutes.
Dentsply Sirona
Operating Instructions inFire HTC speed
61 30 459 D 3497
D 3497.201.01.19.02 09.2017