Regular care
9.1
General care instructions
The high-temperature furnace does not require any special care. The
casing may be cleaned with a mild detergent.
NOTICE
Damage to the heating unit
Avoid all contamination in the heating room to prevent damage to the
heating unit.
NOTICE
Acidic coloring liquids
Coloring liquids may contain acids which can escape during the
sintering process, significantly reducing the service life of the heating
elements.
Recommendation: Use Sirona water-based coloring liquids:
- inCoris TZI coloring liquids.
9.2
Sirona service program
This sintering furnace is equipped with heating elements made from
molybdenum-discilicide. This material is coated with a silicon oxide
protective layer when delivered. The protective layer on the heating
elements prevents interactions with the restorations to be sintered.
The protective layer is normally formed by itself in the event of changing
heating element programs. After each use of the furnace, the protective
layer on the heating element can be worn out.
The service program warms the furnace to an ideal temerpature, at which
the protective layer on the heating elements is fully regenerated.
Using the service program (P12) once a month is recommended.
In the event of sole use of programs 2, 3, and 4, after 30 days, a note
appears on the display requesting the use of the service program. The
display message is hidden for 10s. A previously started program is then
launched.
IMPORTANT
The service program requires approx. 6 h to regenerate the heating
elements. This means that the furnace cannot be used during this time.