5 Setup
5.1 Installation site
14
5

Setup

5.1

Installation site

The unit is designed for desktop use and requires a level footprint of
approx. 500 x 600 mm with a loading capacity of 80 kg.
The unit must not be installed at sites with a high level of humidity or dust!
➢ Always set up the unit in a dry location, as specified in the safety
information.
NOTICE
Danger of overheating
If the unit overheats, the electronics switch off.
➢ Never obstruct the ventilation slots!
➢ Provide ventilation clearance on all sides.
➢ A minimum clearance of 500 mm must be adhered to, above and to
the sides.
Dentsply Sirona
Operating Instructions inFire HTC speed
61 30 459 D 3497
D 3497.201.01.19.02 09.2017