Crosshairs Offset

The laser point and the crosshairs in the Pointfinder do not coincide.
Before Calibration:
a
b
3D Disto_029
a) Crosshairs
b) Laser point
Step
Description
1.
Calibration is started in Menu » Device » Calibration.
2.
Press
.
3.
Pointfinder opens
Place a target mark at a distance >25 m.
4.
Now aim at the target mark as precisely as possible. Press
laser spot is exactly on the target.
Pointfinder stays open and red crosshairs are displayed
5.
Move crosshairs to the centre of target mark as precisely as possible with the
arrow keys. Press
6.
If in tolerance, pop-up appears with Set new: x=...px; y=...px, Reset to
factory default, or Cancel calibration.
7.
Choose Set... or Reset... to adjust crosshairs. Final pop-up asks Are you
sure? Yes/Cancel.
8.
If Yes, checkmark is displayed to confirm parameters were set successfully.

V-Index Error

The vertical index does not coincide with the standing axis.
Tilt sensor calibration followed by a crosshairs calibration is recommended
before starting V-Index calibration! Refer to "8.2 Tilt Sensor Calibration" and
"8.3 Crosshairs Offset".
Before Calibration:
a
b
3D Disto_031
a) Height error
b) Vertical angle offset
www.
After Calibration:
3D Disto_030
.
again.
After Calibration:
3D Disto_032