Replacing The Signal Generator; Replacing The Lid Lock - Hermle Z 326 Service Instruction

Universal centrifuge
Hide thumbs Also See for Z 326:
Table of Contents

Advertisement

• Remove the 2 screws on the backside of the mounting plate and remove the break resistance.
• Put in the new break resistance, connect the electrical connections and re-assemble the unit in reverse
order.

7.4.16 Replacing the signal generator

• Remove the lower covering of the front housing as described in chapter 7.4.1.
• Remove all electrical connections on the control panel (1).
• Go with a flad, sharp item (knife or screwdriver) under the signal transmitter and lift it up with slight
efforts.
• Fix the new signal transmitter with double-sided adhesive tape and re-assemble the unit in reverse
order.

7.4.17 Replacing the lid lock

• Remove the lower covering of the front housing as described in chapter 7.4.1.
• Remove the electrical supply on the control board (1).
• Remove the fixing screws and put in the new lid lock.
Page 37 of 48
1
SERVICE INSTRUCTION Z 326
7 SERVICE INSTRUCTIONS