Imbalance Shut Off Data - Hermle Z 326 Service Instruction

Universal centrifuge
Hide thumbs Also See for Z 326:
Table of Contents

Advertisement

6.3 Imbalance shut off data Z 326 (120 V / 230 V)
List of rotors which cannot be adjusted, but which are set by the work of using motion detectors.
Rotor-Number
220.78 V05
221.18 V02
221.20 V02
221.22 V02
221.28 V02
221.17 V03
220.87 V09
221.87 V10
221.23 V02
220.88 V07
220.92 V06
221.38 V01
List of rotors which can be adjusted by the adjustment knob. See chapter 7.2.9 and 7.4.13.
Rotor-Number
221.12 V03
220.72 V06
221.16 V03
221.19 V02
221.54 V02
221.55 V02
The acceleration and deceleration times may have slight fluctuations that depend on the unit and are
therefore guidelines only.
Page 5 of 48
Shut off speed
Permitted imbalance
in rpm
in gramme
8500
8000
10000
10500
10500
7000
8500
8500
11000
5000
8500
13000
Shut off speed
Permitted imbalance
in gramme
(approx. values) in
rpm
1200
600
1200
1200
800
800
SERVICE INSTRUCTION Z 326
6 TECHNICAL DATA
Imbalance shut off
in gramme
5
5
10
5
5
5
3
3
3
3
3
3
2
Imbalance shut off
in gramme
5
13
14
22
23
11
12
10
11
9
10
© HERMLE Labortechnik GmbH
8
7
9
9
5
5
5
4
5
4
4
6
Version 01/2010