Operation Menu - Hermle Z 326 Service Instruction

Universal centrifuge
Hide thumbs Also See for Z 326:
Table of Contents

Advertisement

A movement sensor, which is mounted in the motor, is controlling the oscillating motion of the motor.
The unit stoops when the oscillating motions are too strong. Setup data specified by the manufacturer for
rotors controlled by a movement sensor cannot be changed.
Break Resistance
on the back side
Chapter 7.2.4

7.3 Operation menu

Valid for the units of the category with frequency converter
(Units: Z20r, Z216MK, Z32HK, Z326K, Z 326, Z36HK, Z366, Seta2)
The operation menu helps the service personnel to locate defects. It is divided in two parts, which are
graded in several areas again from chapter 7.3.1 to 7.3.2.3.
Part 1 is accessible for the USER and different points can be read respectively settled.
This is also described in the instruction manual.
Part 2 is for internal use respectively for trained service personnel outdoors Hermle L
only. The control board can be adjusted to different units. Here the different parameters are determined
on the respective unit after an exchange of the control board.
Page 10 of 48
Chapter 7.2.1
SERVICE INSTRUCTION Z 326
7
SERVICE INSTRUCTIONS
Chapter 7.2.6
Chapter 7.2.9
Chapter 7.2.8
Signal Transmitter
Chapter 7.2.2
ABOR
© HERMLE Labortechnik GmbH
T
ECHNIK
Version 01/2010