7.4 Mounting support

After performing any kind of assembling work, please make sure all the grounded contacts are
connected correctly!
7.4.1 Replacing the front housing, the incremental shaft encoder and the display

Replacing the front housing

• Remove the lower covering by turn out the 2 screws (see photo 19).
• Pull the lower edge of the covering away from the mounting plate.
• Pull the covering out of the upper guiding rail.
• Loosen the 5 screws under the control panel and pull away the panel forward.
• Remove the interface cable on the control board.
(Connection between control board and control panel).
Page 26 of 48
remove
Loosen and
pull away
forward
abziehen
SERVICE INSTRUCTION Z 326
7 SERVICE INSTRUCTIONS
loosen
© HERMLE Labortechnik GmbH