7.3.1.7 Submenu Setup of temperature:
(for refrigerated units only)
Here you can change the temperature unit from degree Celsius (°C) to degree Fahrenheit (°F)..
• Activate the operation menu as described in chapter 7.3.1.
• Turn the potentiometer until the special digit „service c" appears in the display „acc/dec".
• By pressing the key rpm/rcf the display „rpm/rcf" is now activated and you can setup the
respective temperature unit with the potentiometer.
• Now the temperature is indicated in °C resp. in °F during the normal operation.
Attention: After all points have been reviewed and/or set up, you have to press shortly the key
start to store all parameters. Otherwise the setups were all for nothing. If the storage has been
successful in the display „rpm/rcf" the value of the imbalance sensor and the word „StorE" are
indicated.

7.3.1.8 Submenu Setup of the signal generator:

Here you can switch on resp. switch off the signal generator (audible sound).
• Activate the operation menu as described in chapter 7.3.1.
• Turn the potentiometer until the special digit „service L" appears in the display „acc/dec".
• By pressing the key rpm/rcf the display „rpm/rcf" is now activated and you can setup the
respective option with the potentiometer.