Do you have a question about the Z 446 and is the answer not in the manual?
Questions and answers
Dr zekari
February 20, 2025
Bonjour, les tubes commencent a se briser à l intérieur de la centri d une manière aléatoire, l équilibre est respecté, ceci est arrivé après le passage des techniciens pour l étalonnage, sachant qu ils ont vérifié la vitesse, le temps et le froid sans toucher au rotor. Merci
1 comments:
Mr. Anderson
February 20, 2025
Tube breakage inside the Hermle Z 446 centrifuge after calibration, despite maintaining balance, could be caused by several factors:
1. Rotor or Accessory Damage: Even slight damage to rotors, buckets, or tube racks can lead to breakage. Anodized aluminum parts are particularly sensitive.
2. Corrosive Substance Exposure: If rotors, buckets, or tube racks come into contact with corrosive substances like alkalis, alkaline soap solutions, alkaline amines, or concentrated acids, they may weaken and cause breakage.
3. Improper Cleaning and Maintenance: Failing to properly clean and lubricate rotor components, especially aerosol-tight rotor seals and insert bolts, may lead to mechanical stress and tube breakage.
4. Use of Non-Compatible Rotors or Accessories: Only Hermle original rotors and accessories should be used. Non-compatible parts may not function correctly and could cause breakage.
5. Exceeding Sample Load Limits: Overloading or using tubes that are not rated for the centrifuge’s speed or force can result in breakage.
6. Hidden Cracks or Weakness in Tubes: Tubes with pre-existing cracks or defects may break under centrifugal force, even if balance is maintained.
Ensuring proper maintenance, using appropriate accessories, and avoiding corrosive substances can help prevent tube breakage.
For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type Z 446, is a non refrigerated universal centrifuge, which we offer in two voltage variations 230V or 120V. The centrifuge can be used with swing-out rotors and angle rotors.
Instructions for Disposal (see 31) Direction of Rotation – clockwise rotation for the rotor drive Reference for Loading Rotors 1.6.2 Product Nameplate (Example) Company Address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen TYPE: Type Designation of the Product REF: Order No. of the Product Serial No.
EN 61010-2-020. • The HERMLE Z 446, is "non-explosion proof" and must not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. If used in such environment, this is at the users own expense.
Do not run the centrifuge, if mechanical or electrical assembly groups have been tampered with, by unauthorized personnel. • Do not use accessories such as rotors and buckets, that are not approved by HERMLE Labortechnik GmbH, except commercially available centrifuge tubes, made of glass or plastic.
1.7 Installation of the Centrifuge 1.7.1 Unpacking the Centrifuge Model Z 446 is supplied in a carton. Remove the strap retainer, open the carton, and remove the padding. Lift the centrifuge on both sides (see Figure 1) with an appropriate number of helpers and place it on the laboratory table.
In principle, swing out rotors cannot be removed during operation, until all buckets or racks are placed inside the rotor. The bolts of the rotor must be greased with the HERMLE Lubricatinggrease (Order No. 38-5656). The sample tubes have to be filled evenly, by eye, and set into the drillings or tube racks.
Following, lubricate the bolts or sealing, with the recommended HERMLE Rotorgrease - Order No.: 38-5656. Please pay special attention to anodized aluminum parts. Breakage of rotors can be caused by the slightest damages.
"TÜV Nord CERT GmbH, Certification Body Consumer Products, Essen (Germany)", in accordance with Annex AA of IEC 61010-2-020. The certificates can be downloaded on our webpage, www.hermle-labortechnik.de. Aerosol-tight rotors and buckets are marked with the label, "aerosol- tight". ATTENTION: Autoclaving, mechanical stresses and contamination, by chemicals or other aggressive solvents, can impair the aerosol-tightness of the rotors and buckets.
RoHS II Compliance HERMLE Labortechnik GmbH, Siemensstraße 25, 78564 Wehingen, hereby declares and certifies that all components manufactured are RoHS II compliant, according to the definition and restrictions given by the European Parliament Directive 2011/65/EC. This restricts the use of certain hazardous substances in electrical and electronic equipment.
Z 206 A; Z 207 A; Z 207 H, Z 167 M; Z 207 M, Z 216 M; Z 287 A, Z 306; Z 326; Z 366; Z 446; Z 207 MK; Z 216 MK; Z 32 HK; Z 326 K; Z 366 K; Z 36 HK; Z 446 K, ZK 496 Einschlägige EG-Richtlinien / Normen...
Page 43
APPENDIX Table 1: Technical Data HERMLE Labortechnik GmbH, 78564 Wehingen Manufacturer Z 446 Type Dimensions 54 cm Width 67 cm Depth Height 39 cm 79 kg Weight without rotor Max. Speed 16000 min 4 x 750 ml Max. Volume 24325 x g Max.
Page 44
APPENDIX Table 2: Permissible Net Weight Rotor- Max. Speed Number Value 4500 min-1 1420 g 220.86 V20 4500 min-1 2832 g 221.08 V20 4500 min-1 620 g 221.16 V20 15000 min-1 106 g 221.17 V20 11000 min-1 840 g 221.18 V20 15000 min-1 560 g 221.20 V20...
Need help?
Do you have a question about the Z 446 and is the answer not in the manual?
Questions and answers
Bonjour, les tubes commencent a se briser à l intérieur de la centri d une manière aléatoire, l équilibre est respecté, ceci est arrivé après le passage des techniciens pour l étalonnage, sachant qu ils ont vérifié la vitesse, le temps et le froid sans toucher au rotor. Merci
Tube breakage inside the Hermle Z 446 centrifuge after calibration, despite maintaining balance, could be caused by several factors:
1. Rotor or Accessory Damage: Even slight damage to rotors, buckets, or tube racks can lead to breakage. Anodized aluminum parts are particularly sensitive.
2. Corrosive Substance Exposure: If rotors, buckets, or tube racks come into contact with corrosive substances like alkalis, alkaline soap solutions, alkaline amines, or concentrated acids, they may weaken and cause breakage.
3. Improper Cleaning and Maintenance: Failing to properly clean and lubricate rotor components, especially aerosol-tight rotor seals and insert bolts, may lead to mechanical stress and tube breakage.
4. Use of Non-Compatible Rotors or Accessories: Only Hermle original rotors and accessories should be used. Non-compatible parts may not function correctly and could cause breakage.
5. Exceeding Sample Load Limits: Overloading or using tubes that are not rated for the centrifuge’s speed or force can result in breakage.
6. Hidden Cracks or Weakness in Tubes: Tubes with pre-existing cracks or defects may break under centrifugal force, even if balance is maintained.
Ensuring proper maintenance, using appropriate accessories, and avoiding corrosive substances can help prevent tube breakage.
This answer is automatically generated