HERMLE Z 366 K
User Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Z 366 K and is the answer not in the manual?

Questions and answers

Summary of Contents for Hermle Z 366 K

  • Page 1 HERMLE Z 366 K User Manual...
  • Page 3: Table Of Contents

    Pre-Selection .............................21 2.3.1 Pre-Selection of Speed / RCF-Value ....................... 21 2.3.2 Pre-Selection of Running Time ........................ 22 2.3.3 Pre-Selection of Brake Intensity and Acceleration .................. 23 2.3.4 Pre-Selection of Temperature ......................... 24 2.3.5 Pre-Cooling ............................. 24 Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 4 EG - Conformity Declaration ..........................III Table 1: Technical Data ............................ IV Table 2: Permissible Net Weight ........................V Table 3: Lowest Temperature at Max. Speed ....................V Table 4: Max. Speed and RCF-Values for Permissible Rotors ................. VI © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 5 Table 7 (part 1): Radius Correction ......................... VIII Table 7 (part 2): Radius Correction ........................IX Table 7 (part 3): Radius Correction ........................X Table 8: Abbreviations ............................X Redemption Form / Decontamination Certificate ..................... XII Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 6: Product Description

    For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type, Z 366 K, is a refrigerated universal centrifuge, which we offer in two voltage variations 230V or 120V. The centrifuge can be used with swing-out rotors and angle rotors.
  • Page 7: Operating And Display Elements

    Control Panel Display rpm/rcf Speed/ g-force accel/decel Acceleration- / Deceleration Intensity time Centrifugation Time Lid Release quick Short Running start Start Centrifugation stop Stop Centrifugation prog Retrieving Stored Programs store Program Store temp Temperature Indication © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 8: Lcd-Display

    "error" "rpm" "service" "rcf" h m s "accel" "temperature" Indication: After switching on the centrifuge, the display, "rpm/rcf"(A-1) shows the loading status, the current software version and finally the model type (e.g. Z 36 HK). © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 9: Signs And Indications Of The Centrifuge

    Instructions for Disposal (see P. 34) Direction of Rotation – clockwise rotation for the rotor drive Reference for Loading Rotors 1.6.2 Product Nameplate (Example) Company address Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen In-Vitro-Diagnostic Medical Device TYPE: Type Designation of the Product REF: Order No.
  • Page 10: Warning And Information Signs

    EN 61010-2-020. • The HERMLE Z 366 K, is a non "explosion-proof" and must not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. If used in such environment, this is at the users own expense.
  • Page 11: Following Rules Must Strictly Be Adhered To

    Do not run the centrifuge, if mechanical or electrical assembly groups have been tampered with, by unauthorized personnel. • Do not use accessories such as rotors and buckets, that are not approved by HERMLE Labortechnik GmbH, except commercially available centrifuge tubes, made of glass or plastic.
  • Page 12: Installation Of The Centrifuge

    1.7 Installation of the Centrifuge 1.7.1 Unpacking the Centrifuge Model Z 366 K is supplied in a carton. Remove the strap retainer, open the carton, and remove the padding. Lift the centrifuge on both sides (see Figure 1) with an appropriate number of helpers and place it on the laboratory table.
  • Page 13: Attention Before First Use

    If the motor shaft is conical remove the collet from the rotor by turning it The collet is no longer counter clockwise needed If the motor shaft is cylindrical Let the collet inside the rotor © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 14: Basic Adjustments

    All changed settings must be confirmed by the key, "start" (9). A confirmation screen will appear with the word; "store" in the display box, "rpm/rcf"(A-1) - Only then, the pre-selections are valid, (see Figure 3)! Figure 3 © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 15: Temperature Indication

    By pressing the key, "rpm/rcf" (4), the word "On" flashes, and the sound can be switched off with the adjustable knob (1), (see Figure 5). After the settings have been stored by user, the normal program mode can be changed back again by switching off the centrifuge, for a short period. Figure 5 © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 16: Volume Pre-Selection Of Sound Signal

    After pressing the key „rpm/rcf“ (4), select a song with the adjusting knob (1), (see Figure 7). After the settings have been stored by user, the normal program mode can be changed back again by switching off the centrifuge, for a short period. Figure 7 © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 17: Keyboard Sound Turn On / Off

    (1), (see Figure 8). After the settings have been stored by user (see 1.8.1), the normal program mode can be changed back again, by switching off the centrifuge, for a short period. Figure 8 © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 18: Retrieving Operating Data (Operated By Skilled Or Service Engineer Only!)

    (1). The respective error codes appear in the display, "rpm/rcf" (A- 1). Please refer to Table 6: "Error Messages“, (see APPENDIX P.VIII). Switch off the centrifuge, to return to the normal program mode. Figure 9 © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 19: Operation

    Hold the rotor with one hand and secure the rotor to the shaft, by turning the fixing nut clockwise. Tighten the fixing nut with the provided Allen key, (see Figure 11). Figure 12 ATTENTION: For safety, always ensure that the rotor fixing screw is tightened before each run, (see Figure 11)!! © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 20: Loading Angle Rotors

    The bolts of the rotor must be greased with the HERMLE Rotorgrease (Order No. 38-5656). The sample tubes have to be filled evenly, by eye, and set into the drillings or tube racks. The weight difference of the loaded buckets should not exceed 1.0 g.
  • Page 21: Loading And Overloading Of Rotors

    In case of any questions, please contact the manufacturer! 2.1.5 Removing the Rotor Completely, untighten the rotor fixing nut (2. screw over the stiff point) and lift the rotor vertically out of the centrifuge, (see Figures 10 and 11). © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 22: Lid

    For all number marked text, please refer to Figure 17. ATTENTION Don't grip your fingers between the lid and the device or the locking mechanism, when closing the lid! Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 23: Pre-Selection

    See Table 4: "Max. Speed and RCF-Values for Permissible Rotor", (see APPENDIX P. VII). All important values are listed on this table. ATTENTION: Please check the maximum permissible revolutions of your test tubes! (Producer Indication). © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 24: Pre-Selection Of Running Time

    The time counts down, as soon as the set speed is reached. The display will always show the remaining running time, (see Figure 19). For all number marked text, please refer to Figure 19. Figure 19 Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 25: Pre-Selection Of Brake Intensity And Acceleration

    For all number marked text, please refer to Figure 20. See Table 5: "Acceleration and Deceleration Times", (APPENDIX P. VII). This table shows the acceleration and deceleration times, for the acceleration and deceleration stages 0 to 9, for permissible rotors. Figure 20 © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 26: Pre-Selection Of Temperature

    “stop” key (10). Depending on the inserted rotor, the pre-cooling lasts between 10 to 20 min. For all number marked text, please refer to Figure 22. Figure 22 Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 27: Radius Correction

    As soon as the radius correction is set, the word "radius" (M9) appears. This text remains visible until the radius correction is set back to 0. For all number marked text, please refer to Figure 23. Figure 23 © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 28: Program

    (M10) disappears. As soon as the key "store" (12) is no longer displayed, the word "programm xx" (M10) reappears. ( the xx stands for the chosen program place). If all program numbers are occupied, take an old number that is not needed any longer and replace it with the new parameters. Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 29: Recall Of Stored Programs

    To leave the program mode, press the key, "prog" (11). Then, inside the display "time", appears the word "programm". Set the display to "programm--" (M10) with the adjustable knob (1). For all number marked text, please refer to Figure 27. © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 30: Starting And Stopping The Centrifuge

    With the "stop" key (10), the run time can be interrupted, at any time, (see Figure 29). After pressing the key, the centrifuge decelerates with the respective pre-selected intensity, down to a standstill. Figure 29 Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 31: Imbalance Detection

    „rpm/rcf“ (4). The display „time“ (A-3) shows the remaining time until the start of the centrifugation. The timer starts automatically after setting the needed value. The set time can be adjusted by pressing again the bottom “time” (6) and turning the adjusting knob (1). © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 32 You can interrupt this function by pressing the bottom “stop” (10) or the bottom “lid” (11). During the timer mode, all the other bottoms have no function. For all number marked text, please refer to Figure 31 and 32. Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 33: Maintenance

    "TÜV Nord CERT GmbH, Certification Body Consumer Products, Essen (Germany)", in accordance with Annex AA of IEC 61010-2-020. The certificates can be downloaded on our webpage, www.hermle-labortechnik.de. Aerosol-tight rotors and buckets are marked with the label, "aerosol- tight". © Hermle Labortechnik GmbH...
  • Page 34: Cleaning And Disinfecting Of The Unit

    In case of infectious material spilling into the centrifuge, the rotor and rotor chamber have to be disinfected, promptly after the run. Rotors may be autoclaved, the recommended time for autoclaving: 15 – 20 min at 121°C/250°F, (2.15 bar). Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 35: Disinfection Of Pp-Rotors

    Transparent rotor lids and caps, made of PC or PP, as well as rotors, tube racks and adapters of PP, have a maximum operating time of up to 3 years, from first time use. Conditions For the Operating Time: Proper use, damage-free condition, recommended care. © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 36: Trouble Shooting

    • Turn the box spanner to the right (clockwise), up to the limit. ATTENTION: Only turn to the limit, don´t tighten the nut. • Open the lid of the centrifuge. • Switch the centrifuge on again, to proceed with regular function. Figure 33 Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 37: Description Of The Error Message System

    Once the rotor has come to a standstill, open the centrifuge lid with the emergency release. Proceed as described in chapter 4.2.1, p 26. After opening the centrifuge lid, switch on the device again. Error 14 and the lettering "USEr GuidE" should be eliminated. © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 38: Procedure While Error 38 - Lid Motor Is Blocked

    In case of a return of a centrifuge, we also ask you to send its power cord. This eliminates the risk of a faulty power cord. If no power cord is attached to the centrifuge, a new one will be delivered and charged. We ask for your understanding. Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 39: Transport, Storage And Disposal

    The device can be disposed of by stating the date of purchase and the device number at: Hermle Labortechnik GmbH, Siemensstraße 21, 78564 Wehingen, WEEE-Reg. No. DE 55649821 For all devices delivered before 13.08.2005, the last user is responsible for proper disposal.
  • Page 40: Appendix

    Table 7 (part 1): Radius Correction ......................VIII Table 7 (part 2): Radius Correction ......................IX Table 7 (part 3): Radius Correction ......................X Table 8: Abbreviations ..........................X Redemption Form / Decontamination Certificate ................... XII Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 41: Eg - Conformity Declaration

    Z 326; Z 327; Z 366; Z 446; Z 207 MK; Z 216 MK; Z 32 HK; Z 326 K; Z 327 K; Z 366 K; Z 36 HK; Z 446 K; ZK 496 Einschlägige EG-Richtlinien / Normen / Relevant EC Directives / Standards (EU) 2017/746;...
  • Page 42 MAINTENANCE Table 1: Technical Data Manufacturer HERMLE Labortechnik GmbH, 78564 Wehingen Type Z 366 K Dimensions Width 72 cm Depth 51 cm Height 36 cm Weight without rotor 77 kg Max. Speed 16000 rpm Max. Volume 6 x 250 ml Max.
  • Page 43 4500 rpm 2°C/36°F 221.61 V20 16000 rpm 4°C/39°F 221.68 V20 *All temperature indications refer to a room temperature of 23°C/73°F. By exceeding this value or direct solar radiation to the centrifuge, these values cannot be maintained. © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 44 1745 221.20 V20 2906 221.21 V20 1323 221.22 V20 1439 221.28 V20 2395 221.52 V20 221.61 V20 221.68 V20 in seconds Acceleration Time Deceleration Time from 0 rpm -> U from U -> 0 rpm Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 45 Timeout between control unit and frequency converter Other error frequency converter Over Speed Timeout between controller and RS232 interface Rotor is not allowed in this centrifuge FALSE Inserted rotor does not exist in the program rotor no Rotor is not detected © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 46 15.9 628.015 16.3 628.016 15.7 Swing Out Rotor 221.61 16,5 630.000 16.5 630.001 16.5 630.002 15.9 630.003 15.5 630.004 15.6 630.005 15.8 628.016 15.5 Angle Rotor 12.5 221.19 10.9 701.011 12.2 10.6 701.012 10.5 VIII Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 47 707.014 707.000 Angle Rotor 221.22 708.003 708.004 708.017 708.019 Angle Rotor 221.28 Angle Rotor 220.87 704.004 704.005 Angle Rotor 221.52 708.050 12.7 708.051 12.8 701.011 11.4 701.012 701.015 Swing Out Rotor 12.0 221.16 10.1 706.000 © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 48 707.014 10.4 707.015 10.4 Angle Rotor 221.68 704.004 704.005 Table 8: Abbreviations Symbol/Abbreviations Unit Description U (=rpm) [min Revolutions per Minute RZB(=rcf) [x g] Relative Centrifugal Force Polypropylene Polycarbonate accel Acceleration decel Deceleration prog Program Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 49 REPAIR © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 50 Have these substances reached the equipment/assembly? …………………………..  Yes  No If so, which ones: ________________________________________________________________ Description of the measures for the decontamination of the listed parts: I confirm the proper decontamination: Company/Dept ._____________ Place and Date: ____________________________________ Signature of authorized person:_______________________________________ Z366K_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 51 © Hermle Labortechnik GmbH Z366K_V1.22_eng...
  • Page 52 HERMLE Labortechnik GmbH Siemensstraße 25 78564 Wehingen Tel: +49 (0) 74 26 96 22-17 Fax: +49 (0) 74 26 96 22-49 Email: vertrieb@hermleLT.de Internet: http://www.hermle-labortechnik.de Technische Änderungen vorbehalten. ©HERMLE Labortechnik GmbH Z366K_V1.22_eng © Hermle Labortechnik GmbH...