7.3.1.11 Submenu Setup of the signal melody:

Here you can setup the melody at the end of the run in case the signal generator is activated.
• Activate the operation menu as described in chapter 7.3.1.
• Turn the potentiometer until the special digit „service G" appears in the display „acc/dec".
• By pressing the key rpm/rcf the display „rpm/rcf" is now activated and you can setup the
respective option with the potentiometer.
• Now the setup melody will be played by the end of the run, whereas you can choose between
0 and 9 setups (anyway 10 melodies).
Attention: After all points have been reviewed and/or set up, you have to press shortly the key
start to store all parameters. Otherwise the setups were all for nothing. If the storage has been
successful in the display „rpm/rcf" the value of the imbalance sensor and the word „StorE" are
indicated.
7.3.1.12 Submenu Check and calibration of the imbalance sensor (Movement sensor):
Here you can check or re-calibrate the imbalance sensor.
• Activate the operation menu as described in chapter 7.3.1.
• Turn the potentiometer until the special digit „service F" appears in the display „acc/dec".
Non-calibrated indication