1.8.2 Access to the Mode: "Standard settings"

While starting this model, the following basic settings can be performed:
-
-
-
The following operating data can be retrieved in this menu:
-
-
-
-
-
-
-
-
-
-
Open the centrifuge lid and shut off the main switch. Now switch on again the main switch. For
approximately 3 seconds the software version and the model type will be shown in the display (see
Figure 1). Press during this time the keys "time" (6) and "lid/stop" (10) simultaneously. As a result, a
display test is administered for approx. 3 seconds. All possible indications will appear at the same
time (see Figure 3).
Figure 3
ATTENTION:
-
The normal program mode can be changed back again by switching off the centrifuge,
for a short period!
All changed settings must be confirmed by the key "store" or "start" (9). A confirmation
-
screen will appear with the word, "store", in the display "rpm/rcf" (A-1) - Only then the
pre-selections are valid!
1.8.3

Signal Turn On / Off

Proceed as illustrated, under point 1.8.2, to enter this program mode, press the key; "accel/decel"
(5). In the display, "accel/decel" (A-2) flashes the word, "service". Select the letter, "L" with the
control field (1). As a result, appearing in the display "rpm/rcf" (4), are the words, "On Sound". By
© Hermle Labortechnik GmbH
Signal Turn On / Off
Keybord Sound On / Off
Sleep Mode On / Off
Number of Starts
Operating Hours of Centrifuge
Operating Hours of Motor
Software Version
Error List
Function of the Imbalance Switch
Operation of Keyboard
Hardware Version
Intermediate Circuit Voltage in Volt
Display Tests
Z207M_V1.21_eng