HERMLE Z 207 M
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Hermle Z 207 M

  • Page 1 HERMLE Z 207 M User Manual...
  • Page 3: Table Of Contents

    2.4.2 Pre-Selection of Running Time ........................16 2.4.3 Pre-Selection of Brake Intensity and Acceleration..................16 Radius Correction ............................17 Program ................................17 2.6.1 Storage of Programs ............................. 17 2.6.2 Recall of Stored Programs ..........................18 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 4 Table 3: Max. Speed and RCF-Values for Permissible Rotors ................IV Table 4: Acceleration and Deceleration Times ......................IV Table 5: Error Messages ............................IV Table 6: Radius Correction ............................V Table 7: Symbols- / Abbreviations..........................V Redemption Form: Decontamination Certificate ..................... VI Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 5: Product Description

    For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type Z 207 M is a microlitre centrifuge, which we offer in two voltage variations 230V or 120V. The centrifuge can be used with different angle rotors.
  • Page 6: Installation Of The Centrifuge

    Installation of the Centrifuge 1.5.1 Unpacking the Centrifuge Model Z 207 M is supplied in a carton. Remove the strap retainer, open the carton and remove the centrifuge. The instruction manual must be kept with the centrifuge at all times! 1.5.2 Space Requirements...
  • Page 7: Signs- And Indication Of The Centrifuge

    PRODUCT DESCRIPTION 1.6 Signs- and Indication of the Centrifuge 1.6.1 Product Nameplate (Example) Company Address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen TYPE: Type Designation of the Product REF: Order No. of the Product Serial No. of the Product Manufacturer Date of Manufacture MAX.
  • Page 8: Warning And Information Signs

    Attention! Check the fastening of the rotor nut before each run Take off mains plug before opening the housing or the emergency release Power Input Fuse Value 2.5 AT Direction of Rotation – clockwise rotation for the rotor drive Biohazard Warning Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 9: Operating And Display Elements

    Green Run complete, lid is closed Yellow Lid opened, no sleep mode Red flashing fast Error message Red flashing slow Centrifuge runs Yellow flashing slow Sleep mode Red – Green flashing slow Standard settings menu Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 10: Lcd-Display

    „decel“ „open“ „radius“ „rotor“ „programm“ Rotor-Nr. „error“ „rpm“ „service” „rcf“ h m s „accel“ Indication: After switching on the centrifuge, the display "rpm/rcf"(A-1) shows the current software version and after the model type. Figure 1 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 11: Danger, Precautions And Warranty

    EN 61010-2-020. • The HERMLE Z 207 M is non "explosion-proof" and must not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. If used in such environment, this is at the users own expense.
  • Page 12: Warranty

    Figure 2). Select the data in the unit and press the “start” (9) or “store” key. Inside the display, (A-1), the word “store” appears, as the confirmation. Now all important rotor parameters for the centrifuge are stored. Figure 2 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 13: Access To The Mode: "Standard Settings

    Proceed as illustrated, under point 1.8.2, to enter this program mode, press the key; "accel/decel" (5). In the display, "accel/decel" (A-2) flashes the word, "service". Select the letter, "L" with the control field (1). As a result, appearing in the display "rpm/rcf" (4), are the words, "On Sound". By Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 14: Keyboard Sound Turn On / Off

    "SLEEP". After pressing the key, "rpm/rcf" (4), turn the sleep mode function (On) or (Off), with the rotorary field (1). After the settings have been stored by user, the normal program mode can be changed back again by switching off the centrifuge, for a short period. Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 15: Retrieving Operation Data

    (see Figure 4). The first two numbers indicate the appeared errors ongoing from 00 to 99, the last two numbers indicate the error code. Please refer to, Table 5: Error Messages, (see APPENDIX IV). Switch off the centrifuge, to return to the normal program mode. Figure 6 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 16: Operating

    For safety, always ensure that the rotor nut is tightened before each run! Do not operate the centrifuge with rotors or adapters that show any signs of corrosion or mechanical damage. Do not operate with extremely corrosive substances, which could damage the rotor and the centrifuge Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 17: Loading The Angle Rotor

    √ If In case of any questions, please contact the manufacturer! 2.1.4 Removing the Rotor Untighten the rotor nut completely, counter-clockwise. Lift the rotor vertically out of the centrifuge (see Figure 8 and Figure 9). Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 18: Power Switch And Fuse

    "lid/stop" (10), the lid of centrifuge can be released. As soon as the lid is completely released, the word, "open" (M2) appears. The lid of the centrifuge is now able to be opened. For all number marked text, please refer to Figure 13. Figure 13 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 19: Lid Lock

    See Table 3: Max. Speed and RCF-Values for Permissible Rotors, (see APPENDIX P. IV). All important values are listed on this table. ATTENTION: Please notice the maximum permissible revolutions of the test tubes (Producer Indication)! See also chapter 3.1.5 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 20: Pre-Selection Of Running Time

    For all number marked text, please refer to Figure 16. See Table 4: "Acceleration and Deceleration Times", (APPENDIX P. IV). This table shows the acceleration and deceleration times, for the acceleration and deceleration stages 0 and 9, for permissible rotors. Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 21: Radius Correction

    Put the rotor into the centrifuge. By pressing the key, "prog" (11), in the display "time" (A-3) appears the word "program", (see Figure 18). With the control field (1), choose the desired program number. Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 22: Recall Of Stored Programs

    To recall stored programs, press the key "prog" (11), with the lid already closed. Inside the display "time" (A-3), appears "programm --"(M10). With the control field (1), pre-select the desired program number. In the respective displays, the stored values, for that program, will appear. Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 23: Leaving Program Mode

    In the display "time" (A-3), the passed running time is indicated from the moment the "quick" key (8) is pressed. By releasing the "quick" key (8), the centrifuge stops and the running time is indicated, until the lid is opened. For all number marked text, please refer to Figure 22. Figure 22 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 24: The "Stop" Key

    Load the rotor as described in Chapters: 2.1.2 and 2.1.3. When inside the display "time" (A-3), the word "error" along with the number "02" appears, (see Figure 24). Potential reason for this error can be: The imbalance switch is defective. Figure 24 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 25: Maintenance

    The maintenance procedure has to be repeated every 10 to 15 runs, but at least once at week! Connect the unit to the power supply, after the equipment is completely dry. Do not implement disinfection with UV-, beta- and gamma-rays or other high energy radiation. Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 26: Cleaning And Disinfection Of The Unit

    At high temperatures, autoclaving residue may corrode and destroy the plastic. The objects must be thoroughly washed with distilled water, after the cleaning, but before the autoclaving. Residues of any cleaning liquids, may cause fissures, whitening and stains. Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 27: Glass Breakage

    "TÜV Nord CERT GmbH, Certification Body Consumer Products, Essen (Germany)", in accordance with Annex AA of IEC 61010-2-020. The certificates can be downloaded on our webpage, www.hermle-labortechnik.de. Aerosol-tight rotors and buckets are marked with the label, "aerosol- tight". ATTENTION: Autoclaving, mechanical stresses and contamination, by chemicals or other aggressive solvents, can impair the aerosol-tightness of the rotors and buckets.
  • Page 28: Trouble Shooting

    Remove this stopper. Fastened to it, is a string which is connected to the lid lock.. • Pull the string slightly and the lid will open. • Open the lid of the centrifuge. • Switch the centrifuge on again, to proceed with regular function. Figure 25 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 29: Description Of The Error Message System

    4.2.2 Description of the Error Message System The error message, "error" (M11), is shown in the "time" (A-3) display, (see Figure 26). For more detailed information, refer to Table 2: "Error Messages", (see Appendix P. IV). Figure 26 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 30: Receipt Of Centrifuges To Repair

    Health risk from contaminated equipment, rotors and accessories In case of returning the centrifuge for repairing to the manufacturer, please know the following: The centrifuge must be decontaminated and cleaned, before shipment to Hermle Labortechnik, for the protection of persons, environment and material.
  • Page 31: Transport, Storage And Disposal

    RoHS II Compliance HERMLE Labortechnik GmbH, Siemensstraße 25, 78564 Wehingen, hereby declares and certifies that all components manufactured are RoHS II compliant, according to the definition and restrictions given by the European Parliament Directive 2011/65/EC. This restricts the use of certain hazardous substances in electrical and electronic equipmen Z207M_V1.21_eng...
  • Page 32: Appendix

    Table 3: Max. Speed and RCF-Values for Permissible Rotors .........IV Table 4: Acceleration and Deceleration Times ..............IV Table 5: Error Messages .....................IV Table 6: Radius Correction ....................V Table 7: Symbols- / Abbreviations ..................V Redemption Form: Decontamination Certificate ..............VI Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 33: Eg - Conformity Declaration

    Typ Designation Z 206 A; Z 207 A; Z 207 H, Z 167 M; Z 207 M, Z 216 M; Z 287 A, Z 306; Z 326; Z 366; Z 446; Z 207 MK; Z 216 MK; Z 32 HK; Z 326 K; Z 366 K; Z 36 HK; Z 446 K, ZK 496 Einschlägige EG-Richtlinien / Normen...
  • Page 34 APPENDIX Table 1: Technical Data HERMLE Labortechnik GmbH, 78564 Wehingen Manufacturer Z 207 M Type Dimensions 28 cm Width 35 cm Depth 24 cm Height 12.5 kg Weight without rotor Max. Speed 13,500 min Max. Volume 24 x 2…0 ml 17,317 x g Max.
  • Page 35 Imbalance sensor is defective Problem with speed sensor Standstill monitoring defective Wrong dircention of rotoration of the motor Open lid while motor is running Lid contact defective Undervoltage frequency converter Overvoltage frequency converter Error current detection Overspeed Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 36 -1.1 Angle Rotor 221.58 V01 Table 7: Symbols- / Abbreviations Symbol / Unit Describtion Abbreviations n (=rpm) revolutions per minute [min RZB(=rcf) [x g] relative centrifugal force Polypropylen Polycarbonat accel acceleration decel deceleration prog program Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 37: Redemption Form: Decontamination Certificate

    Have these substances reached the equipment/assembly? …………………………..  Yes  No If so, which ones: ________________________________________________________________ Description of the measures for the decontamination of the listed parts: I confirm the proper decontamination: Company/Dept ._____________ Place and Date: ______________________________ Signature of the authorized person: ________________________________________ Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 38 Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 39 NOTES Z207M_V1.21_eng © Hermle Labortechnik GmbH...
  • Page 40 HERMLE Labortechnik GmbH Siemensstraße 25 78564 Wehingen Tel: +49 (0) 74 26-96 22-17 Fax: + +49 (0) 74 26-96 22-49 Email: vertrieb@hermleLT.de Internet: http://www.hermle-labortechnik.de Technical rights reserved ©HERMLE Labortechnik GmbH 2019 Z207M_V1.21_eng © Hermle Labortechnik GmbH...