Cleaning the centrifuge after breakage of glass tubes/bottles ..... 18 E Disinfection ....................18 E 8. 8. 8. 8. 8. Breakdown Breakdown Breakdown Breakdown Breakdown Emergency lid release................... 18 E Check list / Trouble shooting ................. 19 E Z 300 K / July 2000...
For personal and enviromental safety, special attention has to be paid to the following precautions: The Hermle Centrifuges are neither explosion proof nor inert gas shielded and should therefore never be operated in explosion-hazardais locations. Never stay in the safety zone of 30 cm around the centrifuge or deposite dangerous goods inside this zone during centrifugation.
Description The Z 300 K is a refigerated universal centrifuge which covers many fields of applications by offering a wide range of accessories. The centrifuges are equipped with the Hermle Standart-Control panel. Speed, RCF - value, running time and temperature are set with easy to use control knobs.
(e.g. comprising inflammable or infectious liquids) are present material (e.g. comprising inflammable or infectious liquids) are present material (e.g. comprising inflammable or infectious liquids) are present during centrifugation. during centrifugation. during centrifugation. during centrifugation. during centrifugation. Z 300 K / July 2000...
Fill the tubes equally by eye-measuring and insert them into the tube-holes or tube racks. The difference in weight between the buckets should not exceed 10 grams. Incorrect Incorrect Correct Correct Incorrect Incorrect Incorrect Correct Correct Correct sketch 2 Z 300 K / July 2000...
Never operate with strongly corrosive materials which could damage rotor Never operate with strongly corrosive materials which could damage rotor Never operate with strongly corrosive materials which could damage rotor and buckets. and buckets. and buckets. and buckets. and buckets. Z 300 K / July 2000...
The green control lamp of the key "lid" will light as soon as the lid is closed correctly. When the rotor starts accelerating the control lamp of the key "lid" extinguishes and the lid cannot be opened. 10 E Z 300 K / July 2000...
- continuous run Attention: to check the operating time you have to wait one minute after start You can stop a continuous run with key "stop". 12 E Z 300 K / July 2000...
The temperature can be preselected from -10°C up to +40°C. With open lid the temperture display shows the preset temperature and with a closed lid or during a run the display shows the actual rotor temperature. Z 300 K / July 2000 13 E...
If the green control lamp "fast" (2) is lighting, the unit accelerates and decelerates fast. If the green control lamp "slow" (3) is lighting, the unit accelerates and decelerates very softly. slow fast preset speed 2000 14 E Z 300 K / July 2000...
"quick"-key. The operating time will be indicated in seconds on the digital indication "time". Key "stop" Press the key "stop" if you want to interrupt a centrifuge run. The centrifuge decelerates according to the adjusted brake intensity. Z 300 K / July 2000 15 E...
Should the same failure occur again, please call our service. Should the same failure occur again, please call our service. Should the same failure occur again, please call our service. Should the same failure occur again, please call our service. 16 E Z 300 K / July 2000...
By humidity and non hermetically closed samples, condensate (mist) may be formed. The condensate (mist) has to be removed regulary with a cloth from the rotor chamber. Z 300 K / July 2000 17 E...
There is a plastic plug at the left side of the centrifuge housing. Behind that plastic plug there is a red cord. Remove the plastic plug and pull the red cord. The lid can then be opened (see figure 15). figure 15 18 E Z 300 K / July 2000...
Permanent imbalance signal Reason: Position of the imbalance switch not correct Action: Imbalance switch has to be readjusted (call service) Reason: Imbalance switch is defective Action: Imbalance switch has to be replaced (call service) Z 300 K / July 2000 19 E...
Page 20
Start Bit is missing Error no.: 21 Error no.: 21 Start Bit is missing Start Bit is missing Reason: A magnet of the indicator ring is missing Action: Check indicator ring and call service 20 E Z 300 K / July 2000...
Page 21
Error no.: 60 Error no.: 60 Engine speed sensor signal is missing Engine speed sensor signal is missing Reason: Engine speed sensor defective or parting of a cable at the sensor Action: Call service Z 300 K / July 2000 21 E...
Page 22
The power supply is below the tolerance for a short moment Action: Wait till stillstand of the rotor. Open centrifuge lid after the yellow LED "lid" is lighting. Switch off and on the main switch. 22 E Z 300 K / July 2000...