Operating Manual for
Table Top Centrifuge
Z 326

Advertisement

Table of Contents
loading

Summary of Contents for Hermle Z 326

  • Page 1 Operating Manual for Table Top Centrifuge Z 326...
  • Page 3: Table Of Contents

    2.3.2 Pre-Selection of Running Time ........................16 2.3.3 Pre-Selection of Brake Intensity and Acceleration..................17 Radius Correction ............................18 Program ................................18 2.5.1 Storage of Programs ............................. 18 2.5.2 Recall of Stored Programs ..........................19 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 4 Table 6 (Part 1): Radius Correction.......................... X Table 7 (Part 2): Radius Correction......................... XI Table 8 (Part 3): Radius Correction........................XII Table 9: Abbreviations ............................XIII Redemption Form / Decontamination Certificate ....................XV 8. NOTES ..........................Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 5: Product Description

    For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type Z 326, is a non refrigerated universal centrifuge, which we offer in two voltage variations 230V or 120V. The centrifuge can be used with swing-out rotors and angle rotors.
  • Page 6: Operating And Display Elements

    Run Parameters Power Switch Control Panel Display rpm/rcf Speed/ g-force accel/decel Acceleration / Deceleration Intensity time Centrifugation Time Lid Release quick Short Running start Start Centrifugation stop Stop Centrifugation prog Retrieving Programs store Program Store Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 7: Lcd-Display

    "rpm" "service" "rcf" h m s "accel" Signs and Indications of the Centrifuge 1.6.1 General Instructions for Disposal (see P. 28) Direction of Rotation – clockwise rotation for the rotor drive Reference for Loading Rotors Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 8: Product Nameplate (Example)

    PRODUCT DESCRIPTION 1.6.2 Product Nameplate (Example) Company Address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen TYPE: Type Designation of the Product REF: Order No. of the Product Serial Number of the Product Date of Manufacture Manufacturer MAX. Drehzahl: Max. Allowed Speed of the Unit KIN.
  • Page 9: Warning And Information Signs

    EN 61010-2-020. • The HERMLE Z 326, is "non-explosion proof" and must not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. If used in such environment, this is at the users own expense.
  • Page 10: Following Rules Must Strictly Be Adhered To

    Do not run the centrifuge, if mechanical or electrical assembly groups have been tampered with, by unauthorized personnel. • Do not use accessories such as rotors and buckets, that are not approved by HERMLE Labortechnik GmbH, except commercially available centrifuge tubes, made of glass or plastic.
  • Page 11: Installation Of The Centrifuge

    1.7 Installation of the Centrifuge 1.7.1 Unpacking the Centrifuge Model Z 326 is supplied in a carton. Remove the strap retainer, open the carton, and remove the centrifuge. The instruction manual must be kept with the centrifuge, at all times! 1.7.2 Space Requirements...
  • Page 12: Basic Adjustments

    All changed settings must be confirmed by the key, "start" (9). A confirmation screen will appear with the word; "store" in the display box, "rpm/rcf"(A-1) - Only then, the pre-selections are valid, (see Figure 3). Figure 3 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 13: Sound Signal Turn On / Off

    9 (loud), with the adjustable knob (1), (see Figure 5). After the settings have been stored by user (see 1.8.1), the normal program mode can be changed back again by switching off the centrifuge, for a short period. Figure 5 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 14: Song Selection - End Of Run

    (1), (see Figure 7). After the settings have been stored by user (see 1.8.1), the normal program mode can be changed back again by switching off the centrifuge, for a short period. Figure 7 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 15: Retrieving Operating Data

    (1). The respective error codes appear in the display, "rpm/rcf" (A- 1). Please refer to Table 5: "Error Messages“, (see APPENDIX P.IX). Switch off the centrifuge, to return to the normal program mode. Figure 8 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 16: Operation

    Each rotor indicates what the maximum capacity is per hole. (It is allowed to operate e.g. a 12-place- rotor with 2 or 4 loaded tubes only, but the loaded borings must be opposite of each other). Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 17: Loading Swing Out Rotors

    The bolts of the rotor must be greased with the HERMLE Rotorgrease (Order No. 38-5656). The sample tubes have to be filled evenly, by eye, and set into the drillings or tube racks. The weight difference of the loaded buckets should not exceed 1.0 g.
  • Page 18: Loading And Overloading Of Rotors

    In case of any questions, please contact the manufacturer! 2.1.5 Removing the Rotor Completely, untighten the rotor fixing nut (2. screw over the stiff point) and lift the rotor vertically out of the centrifuge, (see Figures 9 and 10). Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 19: Lid

    For all number marked text, please refer to Figure 17. ATTENTION: Don't grip your fingers between the lid and the device or the locking mechanism, when closing the lid! Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 20: Pre-Selection

    3. Range: Continuous Run "cont", can be interrupted by the key, "stop" (10). -The running time can be pre-selected, with the lid opened or closed. -To activate the setting of the running time, press the key "time" (6). Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 21: Pre-Selection Of Brake Intensity And Acceleration

    For all number marked text, please refer to Figure 19. See Table 4: "Acceleration and Deceleration Times", (APPENDIX P. VIII). This table shows the acceleration and deceleration times, for the acceleration and deceleration stages 0 to 9, for permissible rotors. Figure 19 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 22: Radius Correction

    If a program number is already occupied, in the display "rpm/rcf" (A-1), the words "rotor" (M3) and "22x.xx" (M4) will appear, (see Figure 21). Free program numbers will appear as 0. Figure 21 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 23: Recall Of Stored Programs

    "rpm/rcf" (A-1) flashes the word "rotor" (M3). At the same time, the word "FALSE" and the stored rotor number "22x. xx“"(M4) flash in sequence of one another. For all number marked text, please refer to Figure 24. Figure 24 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 24: Leaving Program Mode

    With the "stop" key (10), the run time can be interrupted, at any time, (see Figure 26). After pressing the key, the centrifuge decelerates with the respective pre-selected intensity, down to a standstill. Figure 26 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 25: Imbalance Detection

    Load the rotor as described in Chapters: 2.1.2 and 2.1.3. When inside the display "time" (A-3), the word "error" along with the number "02" appear, (see Figure Potential Reason for the Error Screen: The imbalance switch is defective. Figure 27 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 26: Maintenance

    Please be sure to clean the accessories, especially the sealing of the aerosol- tight rotors and insert bolts, of swing out rotors. Following, lubricate the bolts or sealing, with the recommended HERMLE Rotorgrease - Order No.: 38-5656. Lubricants containing molycote and graphite are not permitted.
  • Page 27: Cleaning And Disinfecting Of The Unit

    "TÜV Nord CERT GmbH, Certification Body Consumer Products, Essen (Germany)", in accordance with Annex AA of IEC 61010-2-020. The certificates can be downloaded on our webpage, www.hermle-labortechnik.de. Aerosol-tight rotors and buckets are marked with the label, "aerosol- tight". ATTENTION: Autoclaving, mechanical stresses and contamination, by chemicals or other aggressive solvents, can impair the aerosol-tightness of the rotors and buckets.
  • Page 28: Disinfection Of Pp-Rotors

    Transparent rotor lids and caps, made of PC or PP, as well as rotors, tube racks and adapters of PP, have a maximum operating time of up to 3 years, from first time use. Conditions For the Operating Time: Proper use, damage-free condition, recommended care. Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 29: Trouble Shooting

    • Turn the box spanner to the right (clockwise), up to the limit. ATTENTION: Only turn to the limit, don´t tighten the nut. • Open the lid of the centrifuge. • Switch the centrifuge on again, to proceed with regular function. Figure 28 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 30: Description Of The Error Message System

    4.2.2 Description of the Error Message System The error message, "error" (M11), is shown in the "time" (A-3) display, (see Figure 29). For more detailed information, refer to Table 5: "Error Messages", (see Appendix P.IX). Figure 29 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 31: Receipt Of Centrifuges To Repair

    Decontamination certificate at goods return delivery, (see APPENDIX P. XV) We reserve the right to accept contaminated centrifuges. Furthermore, all costs that may have occurred during the cleaning and disinfection of the units, will go to the debit of the customer’s account. Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 32: Transport, Storage And Disposal

    RoHS II Compliance HERMLE Labortechnik GmbH, Siemensstraße 25, 78564 Wehingen, hereby declares and certifies that all components manufactured are RoHS II compliant, according to the definition and restrictions given by the European Parliament Directive 2011/65/EC. This restricts the use of certain hazardous substances in electrical and electronic equipment.
  • Page 33: Appendix

    Table 6 (Part 1): Radius Correction ..................X Table 7 (Part 2): Radius Correction ................... XI Table 8 (Part 3): Radius Correction .................. XII Table 9: Abbreviations ..................... XIII Redemption Form / Decontamination Certificate ............XV Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 34: Eg - Conformity Declaration

    Typ Designation Z 206 A; Z 207 A; Z 207 H; Z 207 M; Z 216 M; Z 306; Z 326; Z 366; Z 446; Z 513; Z 216 MK; Z 32 HK; Z 326 K; Z 366 K; Z 36 HK; Z 446 K; ZK 496; Z 513 K Einschlägige EG-Richtlinien / Normen...
  • Page 35 APPENDIX Table 1: Technical Data Manfacturer HERMLE Labortechnik GmbH, 78564 Wehingen Z 326 Type Dimensions Width 40 cm Depth 48 cm Height 36 cm Weight without rotor 43 kg Max. Speed 18000 min Max. Volume 4 x 100 ml Max. RCF...
  • Page 36 220.87 V09 14000 rpm 81.6 g 220.87 V10 14000 rpm 81.6 g 221.23 V02 18000 rpm 40.8 g 220.88 V09 13500 rpm 149.6 g 220.92 V06 13500 rpm 70.4 g 221.38 V01 15000 rpm 14 g Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 37 220.87 V16 14000 rpm 18624 xg 221.23 V02 18000 rpm 23542 xg 220.88 V09 13500 rpm 17113 xg 220.92 V06 13500 rpm 16298 xg 221.38 V01 15000 rpm 15343 xg 221.52 V02 7500 rpm 8174 xg Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 38 221.20 V02 221.22 V02 221.28 V02 220.87 V09 220.87 V10 221.23 V02 220.88 V09 220.92 V06 221.38 V01 in seconds Acceleration Time Deceleration Time from 0 min -> U from U -> 0 min VIII Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 39 Timeout between control unit and frequency converter Other error frequency converter Over Speed Timeout between controller and RS232 interface Rotor is not allowed in this centrifuge FALSE Inserted rotor does not exist in the program rotor no Rotor is not detected Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 40 -0.6 705.016 14.0 -0.6 13.9 -0.7 Angle Rotor 220.87 704.004 -0.4 704.005 -1.1 Angle Rotor 221.54 V02 701.011 10.6 701.012 701.015 Angle Rotor 221.55 V02 708.019 10.7 708.003 10.3 708.004 10.6 701.011 10.2 701.012 701.015 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 41 626.014 -4.9 626.015 11.6 -3.2 Swing Out Rotor 10.2 221.16 706.000 10.2 Angle Rotor 221.17 704.004 -0.4 704.005 -1.1 Angle Rotor 221.20 707.001 -0.7 707.002 -0.8 707.003 -0.3 707.004 -0.6 707.014 -0.9 707.000 -1.7 -0.1 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 42 707.001 10.6 -0.6 707.002 10.4 -0.8 707.003 10.9 -0.3 707.004 10.6 -0.6 707.014 10.4 -0.8 707.015 10.4 -0.8 12.5 Angle Rotor 221.19 10.9 701.000 -4.5 701.010 11.3 -1.2 701.011 12.2 -0.3 10.6 701.012 10.5 -2.0 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 43 APPENDIX Table 9: Abbreviations Symbol / Abbreviation Unit Description U (=rpm) ] Revolutions per Minute [min RZB(=rcf) [x g] Relative Centrifugal Force Polypropylene Polycarbonate accel Acceleration decel Deceleration prog Program XIII Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 44 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 45: Redemption Form / Decontamination Certificate

    Have these substances reached the equipment/assembly? …………………………..  Yes  No If so, which ones: ________________________________________________________________ Description of the measures for the decontamination of the listed parts: I confirm the proper decontamination: Company/Dept ._____________ Place and Date: ____________________________________ Signature of the authorized person: ________________________________________ Z326_V1.16 © Hermle Labortechnik GmbH...
  • Page 46 Z326_V2.16 © Hermle Labortechnik GmbH...
  • Page 47 8. NOTES Z326_V2.16 © Hermle Labortechnik GmbH...