Replacing The Hinges - Hermle Z 326 Service Instruction

Universal centrifuge
Hide thumbs Also See for Z 326:
Table of Contents

Advertisement

7 SERVICE INSTRUCTIONS
7.4.7 Removing the lid
• Remove the lower covering of the front housing as described in chapter 7.4.1.
• Open the lid as wide as possible.
• Hold the lid with one hand and remove the fixing nuts of the hinges.
• The lid can now be removed completely.
fixing nuts

7.4.8 Replacing the hinges

• Remove the lid as described in chapter 7.4.7.
• Lay the lid with the upper side down on a smooth under packing to avoid any scratches.
• Loosen the screws of the respective hinges on the inner side of the lid.
remove
• Re-assemble the unit in reverse order.
• Make sure the lid rests on the lid gasket straight and continuous. Tighten the fasting screws firmly at
the backside of the unit after the adjustment.
Page 30 of 48
SERVICE INSTRUCTION Z 326
© HERMLE Labortechnik GmbH
Version 01/2010