The following table gives information on the individual error messages:
Error No.
Imbalance and rotor error
Temperature error (for refrigerated units only)
General error
Lid error
Converter error
Speed error
Setup error
Page 15 of 48
01
Imbalance occurred
02
Imbalance sensor is defective
04
Imbalance switch has been activated for longer than 5 seconds
from software status 1.28
08
Transponder in the rotor is defective or wrong resp. missing data in
the rotor database
from software status 1.28
11
Temperature sensor is defective
12
Chamber over temperature
14
up to software status 1.32:
Leap of speed is too big between two measurements.
from software status 1.33:
Speed problems; opening of the lid is blocked for a specific time.
33
Lid is open while motor is spinning
34
Lid contact is defective
38
Lid motor is blocked:
all software statuses
from software status 1.26:
the lid lock is blocked for 30 seconds, thereby it can't be overheated
40
Communication with converter disturbed during start
41
Communication with converter disturbed during stop
42
Short circuit in the converter
43
Undervoltage converter
44
Overvoltage converter
45
Over temperature converter
46
Over temperature motor
47
Over current converter
48
Timeout between control unit and converter
49
other error of the converter
55
Overspeed
99
Rotor is not allowed in this centrifuge
SERVICE INSTRUCTION Z 326
7 SERVICE INSTRUCTIONS
Error Description
© HERMLE Labortechnik GmbH
Version 01/2010