HERMLE Z 307
User Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Z 307 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hermle Z 307

  • Page 1 HERMLE Z 307 User Manual...
  • Page 3: Table Of Contents

    Lid Lock ................................... 16 Pre-Selection ..................................17 2.3.1 Pre-Selection of Speed / RCF-Value .......................... 17 2.3.2 Pre-Selection of Running Time............................ 17 2.3.3 Pre-Selection of Brake Intensity and Acceleration ....................18 Radius Correction ................................19 Program....................................19 © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 4 Table 4: Max. Speed and RCF-Values for Permissible Rotors..................V Table 5: Acceleration and Deceleration Times ........................V Table 6: Error Messages ................................. VI Table 7: Radius Correction ..............................VII Table 7(Part 2): Radius Correction ............................VIII Redemption Form / Decontamination Certificate ........................ IX Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 5 CONTENTS © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 7: Product Description

    For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type Z 307, is a non-refrigerated universal centrifuge, which we offer in two voltage variations 230V or 120V. The centrifuge can be used with swing-out rotors and angle rotors.
  • Page 8: Installation Of The Centrifuge

    Installation of the Centrifuge 1.5.1 Unpacking the Centrifuge Model Z 307 is supplied in a carton. Remove the strap retainer, open the carton, and remove the padding. Lift the centrifuge on both sides (see Figure 1) with an appropriate number of helpers and place it on the laboratory table.
  • Page 9: Signs And Indications Of The Centrifuge

    1.6.1 General Direction of Rotation – clockwise rotation for the rotor drive Reference for Loading Rotors 1.6.2 Product Nameplate (Example) Company Address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen In-Vitro-Diagnostic Medical Device TYPE: Type Designation of the Product REF: Order No. of the Product Serial No.
  • Page 10: Warning And Information Signs

    EN 61010-2-020. • The HERMLE Z 307, is "non-explosion proof" and must not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. If used in such environment, this is at the users own expense.
  • Page 11: Following Rules Must Strictly Be Adhered To

    Do not run the centrifuge, if mechanical or electrical assembly groups have been tampered with, by unauthorized personnel. • Do not use accessories such as rotors and buckets, that are not approved by HERMLE Labortechnik GmbH, except commercially available centrifuge tubes, made of glass or plastic. •...
  • Page 12: Operating And Display Elements

    Green Run complete, lid is closed Yellow Lid opened, no sleep mode Red flashing fast Error message Red flashing slow Centrifuge runs Yellow flashing slow Sleep mode Red – Green flashing slow Standard settings menu Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 13: Lcd-Display

    "rpm" "service" "rcf" h m s "accel" Indication: After switching on the centrifuge, the display, "rpm/rcf"(A-1) shows the loading status, the current software version and finally the model type (e.g. Z 36 HK). Figure 4 © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 14: Basic Adjustments

    All changed settings must be confirmed by the key “store” (12) or “start” (9). A confirmation screen will appear with the word, “store”, in the display “rpm/rcf” (A-1), - Only then the pre- selections are valid (see Figure 6). Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 15: Sound Signal Turn On / Off

    "rpm/rcf" (4), the desired volume can be adjusted between 0 (low) and 9 (loud), with the control field (1), (see Figure 8). After the settings have been stored by user, the normal program mode can be changed back again by switching off the centrifuge, for a short period. © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 16: Song Selection - End Of Run

    "rpm/rcf" (4), the keyboard sound (On) or (Off) can be turned on, with the control field (1), (see Figure 10). After the settings have been stored by user, the normal program mode can be changed back again by switching off the centrifuge, for a short period. Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 17: Setting Of The Standby Turn-Off Time

    „rpm/rcf“, the standby mode can be turned off or pre-set to the needed value by turning the control field (1). After the settings have been stored by user, the normal program mode can be changed back again by switching off the centrifuge, for a short period. Figure 11 © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 18: Retrieving Operating Data

    Figure 12 While retrieving the number of rotor starts and operating hours of the mounted rotor, as a result, appearing in the display “rpm/rcf” (A-1), the word “r-StA” respectively “r-hou” (see Figure 13). Figure 13 Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 19: Operation

    Therefore, we recommend to weigh with a balance. This reduces the wear of drive and the acoustic operating noise. Each rotor indicates what the maximum capacity is per hole. © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 20: Loading Swing Out Rotors

    The bolts of the rotor must be greased with the HERMLE Rotorgrease (Order No. 38-5656). The sample tubes have to be filled evenly, by eye, and set into the drillings or tube adapter. The weight difference of the loaded buckets should not exceed 1.0 g.
  • Page 21: Loading And Overloading Of Rotors

    In case of any questions, please contact the manufacturer! 2.1.5 Removing the Rotor Completely, untighten the rotor fixing nut (2. screw over the stiff point) and lift the rotor vertically out of the centrifuge, (see Figures 15 and 16). © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 22: Lid

    "nr 220.72“ (M4), along with all rotor specific data, for example: max. speed, acceleration etc., are available. For all number marked text, please refer to Figure 21. ATTENTION: Don't grip your fingers between the lid and the device or the locking mechanism, when closing the lid! Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 23: Pre-Selection

    To set the desired value, use the control field (1). After exceeding 59 min 50 sec, the indication changes automatically to, "h : m". After exceeding 99 hours 59 min, the word "cont" appears in the © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 24: Pre-Selection Of Brake Intensity And Acceleration

    For all number marked text, please refer to Figure 24. See “Table 5: Acceleration and Deceleration Times", (APPENDIX P. V). This table shows the acceleration and deceleration times, for the acceleration and deceleration stages 0 to 9, for permissible rotors. Figure 24 Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 25: Radius Correction

    "program", (see Figure 26). With the control field (1), choose the desired program number. If a program number is already occupied, in the display "rpm/rcf" (A-1), the words "rotor" (M3) and "22x.xx" (M4) will appear, (see Figure 26). Free program numbers will appear as 0. Figure 26 © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 26: Recall Of Stored Programs

    To leave the program mode, press the key, "prog" (11). Then, inside the display "time", appears the word "programm". Set the display to "program - -" (M10) with the control field (1). For all number marked text, please refer to Figure 29. Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 27: Starting And Stopping The Centrifuge

    2.6.2 The "STOP" Key With the "stop" key (10), the run time can be interrupted, at any time, (see Figure 31). After pressing the key, the centrifuge decelerates with the respective pre-selected intensity, down to a standstill. © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 28: Imbalance Detection

    99 h 59 min by pressing the bottom “time” (6) and turning the control field (1). Alternatively to the “End in” mode, you can choose the “Start in” mode by pressing the bottom „rpm/rcf“ (4). The display „time“ (A-3) shows the remaining time until the start of the centrifugation. Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 29 You can interrupt this function by pressing the bottom “stop” (10) or the bottom “lid” (7). During the timer mode, all the other bottoms have no function. For all number marked text, please refer to Figure 33 and 34. © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 30: Maintenance

    Following, lubricate the bolts or sealing, with the recommended HERMLE Rotorgrease - Order No.: 38-5656. Please pay special attention to anodized aluminum parts. Breakage of rotors can be caused by the slightest damages.
  • Page 31: Cleaning And Disinfecting Of The Unit

    Bottles, adapters and rotors may be treated, with the usual liquid disinfectants. ATTENTION: Before applying any other, Cleaning Resp. Decontamination Method, other than what was recommended by the manufacturer, contact the manufacturer to ensure that it will not damage the unit or the rotor. © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 32: Glass Breakage

    The aerosol tightness of rotors, rotor lids, buckets and caps have been tested and certified by the "TÜV Nord CERT GmbH, Certification Body Consumer Products, Essen (Germany)", in accordance with Annex AA of IEC 61010-2-020. The certificates can be downloaded on our webpage, www.hermle-labortechnik.de. Aerosol-tight rotors and buckets are marked with the label, "aerosol-tight".
  • Page 33: Trouble Shooting

    At the right side of the centrifuge, there is a plastic stopper (Figure 35). Remove this stopper, which is connected to the lid lock, horizontally from the housing until the centrifuge lid opens. • Now, open the lid of the centrifuge. Figure 35 © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 34: Description Of The Error Message System

    Once the rotor has come to a standstill, open the centrifuge lid with the emergency release. Proceed as described in chapter 4.2.1, p 27. After opening the centrifuge lid, switch on the device again. Error 14 and the lettering "USEr GuidE" should be eliminated. Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 35: Procedure While Error 38

    Proceed as described in chapter 4.2.1. After opening the centrifuge lid, switch on again the device. The closing process can be carried out normally again. If these error occurs again, please contact the service. © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 36: Receipt Of Centrifuges For Repair

    In case of a return of a centrifuge, we also ask you to send its power cord. This eliminates the risk of a faulty power cord. If no power cord is attached to the centrifuge, a new one will be delivered and charged. We ask for your understanding. Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 37: Transport, Storage And Disposal

    The device can be disposed of by stating the date of purchase and the device number at: Hermle Labortechnik GmbH, Siemensstraße 21, 78564 Wehingen, WEEE-Reg. No. DE 55649821 For all devices delivered before 13.08.2005, the last user is responsible for proper disposal.
  • Page 39: Appendix

    Table 4: Max. Speed and RCF-Values for Permissible Rotors ..........V Table 5: Acceleration and Deceleration Times ............... V Table 6: Error Messages ......................VI Table 7: Radius Correction ....................... VII Table 7(Part 2): Radius Correction..................VIII Redemption Form / Decontamination Certificate ..............IX © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 40: Ec - Conformity Declaration

    Typenbezeichnung / Typ Designation Z 206 A; Z 207 A; Z 207 H, Z 167 M; Z 207 M, Z 216 M; Z 287 A, Z 306; Z 307; Z 326; Z 327; Z 366; Z 446; Z 207 MK; Z 216 MK; Z 32 HK; Z 326 K; Z 327 K;...
  • Page 41 APPENDIX Table 1: Technical Data Manfacturer HERMLE Laborte c hnik GmbH, 78564 Wehinge n Type Z 307 Dimensions Width 35.5 cm Depth 47.4 cm Height 33.4 cm Weight withou t rotor 30 kg Max. Speed 14,000 min Max. Volume 4 x 200 ml Max.
  • Page 42 6 x 72 g 220.87 V22 14,000 rpm 24 x 3.4 g 220.87 V20 14,000 rpm 24 x 3.4 g 221.17 V21 12,000 rpm 30 x 3.4 g 221.35 V20 14,000 rpm 12 x 9.5 g Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 43 221.12 V20 221.16 V20 220.50 V20 221.19 V20 221.54 V20 221.55 V20 220.87 V22 220.87 V20 221.17 V21 221.35 V20 in seconds Acceleration Time Deceleration Time from 0 rpm -> U from U -> 0 rpm © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 44 Timeout between control unit and frequency converter Other error frequency converter Over Speed Rotor is not allow ed in this centrifuge FALSE Inserted rotor does not exist in the program rotor no Rotor is not detected Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 45 626.006 14.1 626.007 13.9 626.008 14.1 626.009 14.1 626.010 14.2 626.011 13.7 626.012 14.3 626.013 14.4 626.014 626.015 11.5 Swing Out Rotor 12.0 221.16 V20 706.000 12.0 Swing Out Rotor 10.4 220.50 V20 706.000 10.4 © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 46 708.019 10.7 708.020 708.003 10.3 708.004 10.6 708.030 10.7 701.011 701.012 701.015 Fixed Angle Rotor 220.87 V20/22 704.004 704.005 Fixed Angle Rotor 221.17 V21 704.004 704.005 Fixed Angle Rotor 221.35 V20 701.015 701.016 701.017 VIII Z307_V1.22_eng © Hermle Labortechnik GmbH...
  • Page 47: Redemption Form / Decontamination Certificate

    If so, which ones: ________________________________________________________________ Description of the measures f or the decontamination of the listed parts : I conf irm the proper decontamination: Company/Dept ._____________ Place and Date: ____________________________________ Signature of the authorized person: ________________________________________ © Hermle Labortechnik GmbH Z307_V1.22_eng...
  • Page 48 HERMLE Labortechnik GmbH Siemensstraße 25 78564 Wehingen Tel: 0 74 26-96 22-17 Fax: 0 74 26-96 22-49 Email: info@hermleLT.de Internet: http://www.hermle-labortechnik.de Technical modification rights reserved. ©HERMLE Labortechnik GmbH 2022 Z307_V1.22_eng © Hermle Labortechnik GmbH...