For any resulting damage, the company, Hermle Labortechnik, is not liable. 1.3 Brief Description The unit type Z 307, is a non-refrigerated universal centrifuge, which we offer in two voltage variations 230V or 120V. The centrifuge can be used with swing-out rotors and angle rotors.
Installation of the Centrifuge 1.5.1 Unpacking the Centrifuge Model Z 307 is supplied in a carton. Remove the strap retainer, open the carton, and remove the padding. Lift the centrifuge on both sides (see Figure 1) with an appropriate number of helpers and place it on the laboratory table.
1.6.1 General Direction of Rotation – clockwise rotation for the rotor drive Reference for Loading Rotors 1.6.2 Product Nameplate (Example) Company Address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen In-Vitro-Diagnostic Medical Device TYPE: Type Designation of the Product REF: Order No. of the Product Serial No.
EN 61010-2-020. • The HERMLE Z 307, is "non-explosion proof" and must not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. If used in such environment, this is at the users own expense.
Do not run the centrifuge, if mechanical or electrical assembly groups have been tampered with, by unauthorized personnel. • Do not use accessories such as rotors and buckets, that are not approved by HERMLE Labortechnik GmbH, except commercially available centrifuge tubes, made of glass or plastic. •...
The bolts of the rotor must be greased with the HERMLE Rotorgrease (Order No. 38-5656). The sample tubes have to be filled evenly, by eye, and set into the drillings or tube adapter. The weight difference of the loaded buckets should not exceed 1.0 g.
Following, lubricate the bolts or sealing, with the recommended HERMLE Rotorgrease - Order No.: 38-5656. Please pay special attention to anodized aluminum parts. Breakage of rotors can be caused by the slightest damages.
The aerosol tightness of rotors, rotor lids, buckets and caps have been tested and certified by the "TÜV Nord CERT GmbH, Certification Body Consumer Products, Essen (Germany)", in accordance with Annex AA of IEC 61010-2-020. The certificates can be downloaded on our webpage, www.hermle-labortechnik.de. Aerosol-tight rotors and buckets are marked with the label, "aerosol-tight".
The device can be disposed of by stating the date of purchase and the device number at: Hermle Labortechnik GmbH, Siemensstraße 21, 78564 Wehingen, WEEE-Reg. No. DE 55649821 For all devices delivered before 13.08.2005, the last user is responsible for proper disposal.
Typenbezeichnung / Typ Designation Z 206 A; Z 207 A; Z 207 H, Z 167 M; Z 207 M, Z 216 M; Z 287 A, Z 306; Z 307; Z 326; Z 327; Z 366; Z 446; Z 207 MK; Z 216 MK; Z 32 HK; Z 326 K; Z 327 K;...
Page 41
APPENDIX Table 1: Technical Data Manfacturer HERMLE Laborte c hnik GmbH, 78564 Wehinge n Type Z 307 Dimensions Width 35.5 cm Depth 47.4 cm Height 33.4 cm Weight withou t rotor 30 kg Max. Speed 14,000 min Max. Volume 4 x 200 ml Max.