• Remove the screws and the center screw of the electronic components mounting plate. Upon
there are fixed the electronically components.
• Tilt the mounting plate forwards.
• Remove the electrical connections of the break resistance to the frequency converter.
• Remove the fixing screw.
Figure 58
• Put in the new break resistance, connect the electrical connections and reassemble the unit in
reverse order.

9.4.16 Replacing the signal generator

• Remove the covering of the front housing as described in chapter 9.4.1.
• Remove electrical connections on the control panel (see figure 59,1)
Figure 59
• Go with a flat, sharp item (knife or screwdriver) under the signal transmitter and lift it up with
slight efforts.
© Hermle Labortechnik GmbH
remove
Service Manual Z 306
SERVICE INSTRUCTIONS
1
37