PMC PROCEDURE (continued)
6.8.7.3.7 Squeeze the breathing bag attached to the bag mount at a
( )
6.8.7.3.8 Observe the operation of each unidirectional valve disc at eye
6.8.7.3.9 Remove the breathing circuit.
6.8.7.3.10
( ) 6.8.8

Absorber PEEP Valve w/By-Pass - if applicable

6.8.8.1
*
6.8.8.2
6.8.8.3
6.8.8.4
6.8.8.5
*
6.8.8.6
6.8.8.7
6.8.8.8
6.8.8.9
* These items apply only to machines with a PEEP by-pass.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
rate of approximately 10 BPM. Readjust the APL valve if
required to properly ventilate the simulated lung.
level and make sure the inspiratory valve disc raises only
during the inspiration phase, and the expiratory valve raises
only during the exhalation phase. Watch the valves until
satisfied that both valves operate correctly, and move freely
without sticking.
Open the APL valve.
Connect a 12 in. hose between the inspiratory and expiratory
valves, and set the O
2
Place the PEEP bypass in the ON position.
Adjust the absorber PEEP valve clockwise to the maximum
position.
Does the PEEP valve adjust smoothly? ___ (Y)
What is the maximum PEEP? ___ cm H
Place the PEEP bypass in the OFF position.
Does the PEEP return to ≤3 cm H
Adjust the absorber PEEP valve counterclockwise to its minimum
position.
Close the O
flow control valve.
2
6-20
RETURN TO CD-ROM TABLE OF CONTENTS
flow to 5 l/min.
O (15-22)
2
O? ___(Y)
2
NM2B
Rev. Y