Air High Pressure Leak - If Applicable; Oxygen-Helium High Pressure Leak - If Applicable; Carbon Dioxide High Pressure Leak - If Applicable - Dräger Narkomed 2B Service Manual

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

NM2B
( ) 6.13.3

Air High Pressure Leak - if applicable

*
6.13.3.1
6.13.3.2
6.13.3.3
6.13.3.4
6.13.3.5
6.13.3.6
6.13.3.7
( ) 6.13.4

Oxygen-Helium High Pressure Leak - if applicable

*
6.13.4.1
6.13.4.2
6.13.4.3
6.13.4.4
6.13.4.5
6.13.4.6
6.13.4.7
( ) 6.13.5

Carbon Dioxide High Pressure Leak - if applicable

*
6.13.5.1
6.13.5.2
6.13.5.3
6.13.5.4
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
Rev. Y
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Set the gas selector switch to ALL GASES.
Open the air cylinder valve.
Let the pressure stabilize.
Close the air cylinder valve and remove the cylinder.
Observe the air cylinder pressure gauge.
After two (2) minutes, what is the pressure loss? ___ PSI (<50)
Attach the air cylinder.
Set the gas selector switch to ALL GASES.
Open the oxygen-helium cylinder valve.
Let the pressure stabilize.
Close the oxygen-helium cylinder valve and remove the cylinder.
Observe the oxygen-helium cylinder pressure gauge.
After two (2) minutes, what is the pressure loss? ___ PSI (<50)
Attach the oxygen-helium cylinder.
Set the gas selector switch to ALL GASES.
Open the carbon dioxide cylinder valve.
Let the pressure stabilize.
Close the carbon dioxide cylinder valve and remove the cylinder.
6-29
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)