Carbon Dioxide O.f.p.d. - If Applicable; Nitrogen O.f.p.d. - If Applicable - Dräger Narkomed 2B Service Manual

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

NM2B
6.14.4

Carbon Dioxide O.F.P.D. - if applicable

*
6.14.4.1
6.14.4.2
6.14.4.3
6.14.4.4
6.14.4.5
( )
6.14.4.6
6.14.4.7
*
6.14.4.8
6.14.5

Nitrogen O.F.P.D. - if applicable

6.14.5.1
6.14.5.2
6.14.5.3
6.14.5.4
6.14.5.5
( )
6.14.5.6
6.14.5.7
6.14.5.8
6.14.5.9
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
Rev. Y
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Set the gas selector switch to ALL GASES.
Open the cylinder of carbon dioxide.
Open one (1) oxygen cylinder valve.
Set the carbon dioxide flow to 500 ml; set the oxygen flow to 4
l/min.
Close the oxygen cylinder valve.
Does the flow of carbon dioxide cease when the oxygen pressure is
depleted? ___ (Y)
Close the carbon dioxide flow control valve.
Set the gas selector switch to O
Set the gas selector switch to ALL GASES.
Open the nitrogen cylinder valve.
Open one (1) oxygen cylinder valve.
Set the nitrogen flow to 4 l/min; set the oxygen flow to 4 l/min.
Close the oxygen cylinder valve.
Does the flow of nitrogen cease when the oxygen pressure is
depleted? ___ (Y)
Close the nitrogen flow control valve.
Set the gas selector switch to O
Close the oxygen flow control valve.
6-33
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)
+N
O.
2
2
+N
O.
2
2