Baromed - Dräger Narkomed 2B Service Manual

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

NM2B

6.20 BAROMED

6.20.1
Disconnect the breathing pressure sensor line from the absorber.
6.20.2
Connect a test pressure gauge and syringe to the breathing pressure
sensor line.
6.20.3
Press the THRESHOLD key on the main keypad. A box shall appear
around the Threshold Alarm Limit.
6.20.4
What is the threshold alarm default? ___ cm H
6.20.5
Verify that the threshold alarm limit has a range from 5 to 30 cm H
6.20.6
Adjust the threshold to 10 cm H2O.
6.20.7
Press the HIGH key in the Baromed section of the main keypad. A box
shall appear around the High Pressure Alarm Limit.
6.20.8
What is the high alarm limit default? ___ cm H
6.20.9
Verify that the high alarm limit has an adjustable range from 30 to 120
cm H
6.20.10
Set the high alarm limit to 65 cm H
6.20.11
Increase the pressure to 25 cm H
H
6.20.12
Does the THRESHOLD LOW message appear in the Advisory column?
___ (Y)
6.20.13
Set the Man/Auto valve to AUTO, and turn the ventilator ON.
6.20.14
Bleed the pressure and start a stopwatch.
( )
6.20.15
What is the time when APNEA-PRESSURE appears in the Caution
column? ___ sec (13-17)
( )
6.20.16
What is the time when the APNEA-PRESSURE appears in the Warning
column? ___ sec (26-34)
6.20.17
After the APNEA-PRESSURE alarm is displayed in the Warning
column on remote display, slowly increase the test pressure.
( )
6.20.18
At what pressure does the APNEA-PRESSURE alarm deactivate? ___
cm H
Rev. Y
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
O. (Later software versions allow range up to 125 cm H
2
O.
2
O (7-13)
2
6-47
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)
O (12)
2
O (50)
2
O.
2
O; decrease the pressure to 20 cm
2
O.
2
O).
2

Hide quick links: