Dräger Narkomed 2B Service Manual page 243

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

NM2B
7.11 Turn the System Power switch to ON, and observe the software version number
that appears on the power-up diagnostics screen. Verify that it matches the label
on the replacement firmware chip.
7.12 Turn the System Power switch to STANDBY.
7.13 Place the monitor box top shelf into position, connect its ground wire to the chassis
ground tab, and reinstall the top shelf screws.
7.14 Perform the PMS Procedure given in Section 6.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SOFTWARE UPDATE PROCEDURE (continued)
7-3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents