PMC PROCEDURE (continued)

( ) 6.2 SELF-DIAGNOSTICS

6.2.1
Connect the pipeline supply or open the cylinders.
6.2.2
Turn the System Power switch to ON.
6.2.3
Verify that the following is displayed on the alarm CRT:
VIDEO TEST
FIRMWARE TEST
MEMORY TEST
TIMERS TEST
ANALOG TEST
AUDIO TEST - PRIMARY
SERIAL I/O TEST
CLOCK TEST
BACKUP MEMORY TEST
AC POWER TEST
RESERVE POWER TEST
FUNCTIONAL
( )
6.2.4
Record the machine software version on the header of the checklist
form.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
PASS
PASS
PASS
PASS
PASS
PASS
- BACKUP
PASS
PASS
PASS
PASS
PASS
PASS
6-4
RETURN TO CD-ROM TABLE OF CONTENTS
NARKOMED 2B
VERSION x.xx DIAGNOSTICS
COPYRIGHT, NAD INC. 1987-94
NM2B