ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
5.4.31
The nitrous oxide flow should stop when the flow of oxygen is
between 200 and 300 ml per minute.
*
5.4.32
If the O
flow drops to between 100 and 200 ml per minute, proceed to Step
5.4.34.
*
5.4.33
If the O
oxide flow drops to between 100 and 200 ml per minute, adjust the
MPL switch until the O
5.4.34
Close the N
STANDBY.
5.4.35
Replace the rear cover of the vapor box.
5.4.36
Perform the PMS Procedure given in Section 6.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
/N
O FLOW RATIO lamp turns off when the nitrous oxide
2
2
/N
O FLOW RATIO lamp does not turn off when the nitrous
2
2
/N
O FLOW RATIO lamp turns off.
2
2
O flow control valve and set the System Power switch to
2
5-14