Dräger Narkomed 2B Service Manual page 87

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

NM2B
4.17.6
4.17.7
4.17.8
4.17.9
4.17.10 Remove the prop from the rear of the monitor box, and return the box
4.17.11 Replace the screws holding the monitor box to the ventilator box.
4.17.12 Restore power to the machine and perform the PMS Procedure given
Rev. A
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued)
Unscrew the two hex post nuts holding the filter to the ventilator box,
and remove the filter.
Install the replacement filter and secure it with the two hex post
nuts.
Connect the AC wiring to the replacement filter. Be sure to observe
the wire color code as illustrated.
Place the safety cover over the filter and secure it with the two
screws.
to its normal position.
in Section 6.
4-55