Advertisement

PMC PROCEDURE (continued)

6.16 O2 MED

6.16.1
Disconnect the O
6.16.2
The following messages shall appear on the display: "O
"O
2
6.16.3
Reconnect the O
6.16.4
The "O
6.16.5
Press the "21%" key to calibrate the O
NOTE:
Make sure the sensor is stabilized in ambient air for several minutes.
( )
6.16.6
After the calibration is completed, what is the oxygen concentration?
___ % (21)
6.16.7
This step intentionally left blank.
6.16.8
The alarm CRT shall display "% OXYGEN LOW" under the warning
heading. There should be a continuous audible alarm.
6.16.9
What is the low oxygen alarm default? ___ % (30)
6.16.10
This step intentionally left blank.
6.16.11
Press the LO oxygen alarm key. Does a box appear around the low
alarm limit? ___ (Y)
6.16.12
Verify that the low alarm limit has a range from 21 to 99% (18 to 99%
with firmware version 2.03 and later).
6.16.13
Place the sensor into the inspiratory valve dome adapter. Set the
Man/Auto selector to BAG, close the APL valve, attach a 12-inch hose
to the inspiratory valve, and occlude the bag mount.
6.16.14
Set the oxygen flow to 4 l/min.
6.16.15
Set the low limit to the lowest setting possible, and verify the "%
OXYGEN LOW" message has cleared.
6.16.16
Press the HI oxygen alarm key. Does a box appear around the high
alarm limit? ___(Y)
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
sensor cable from the oxygen sensor interface.
2
NOT CAL", "O
ALRM OFF".
2
Med sensor.
2
SENS DISC" message shall disappear.
2
6-38
RETURN TO CD-ROM TABLE OF CONTENTS
Med.
2
NM2B
SENS DISC",
2
Rev. Y