PMC PROCEDURE (continued)
6.14.6

Helium O.F.P.D. - if applicable

6.14.6.1
6.14.6.2
6.14.6.3
6.14.6.4
6.14.6.5
( )
6.14.6.6
6.14.6.7
6.14.6.8
6.14.6.9
6.14.7

Oxygen Supply Pressure Alarm

6.14.7.1
6.14.7.2
6.14.7.3
( )
6.14.7.4
6.14.7.5
6.14.7.6
6.14.7.7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Set the gas selector switch to ALL GASES.
Open the helium cylinder valve.
Open one (1) oxygen cylinder valve.
Set the helium flow to 4 l/min; set the oxygen flow to 4 l/min.
Close the oxygen cylinder valve.
Does the flow of helium cease when the oxygen pressure is
depleted? ___ (Y)
Close the helium flow control valve.
Set the gas selector switch to O
Close the oxygen flow control valve.
If not already connected, connect a dedicated O
TEE fitting in the O
2
Open and close an oxygen cylinder.
Set the oxygen flow to 2 l/min.
What is the pressure on the dedicated O
SUPPLY PRESSURE" LED turns on? ___ PSI (34 - 40)
Does the O2 SUPPLY LOW message appear in the Caution
display? ___ (Y)
Bleed the remaining O
flow control valve.
Remove the test gauge from the TEE fitting in the O
output line and replace the plug.
6-34
RETURN TO CD-ROM TABLE OF CONTENTS
+ N
O.
2
2
regulator output line.
test gauge when the "O2
2
pressure from the system, then close the
2
NM2B
test gauge to the
2
regulator
2
Rev. Y