NM2B
For example:
A Narkomed 3 could have recommendation number 21 and failure code 61.1 apply.
21 - No ventilator pressure limit control. Code is B.
61.1 - Enflurane agent is unavailable to test. Code is BC.
Correct certification for this machine is BC, which means CONDITIONALLY CERTIFIED WITH
RECOMMENDATIONS.
A Narkomed 4 could have recommendation numbers 14 and 21 apply.
14 - CO2/Agent monitor exhaust port is not properly connected to the waste gas scavenger.
Code B.
21 - No ventilator pressure limit control. Code B.
The
correct
certification
RECOMMENDATIONS.
A Narkomed 2B, 2C or GS could have recommendation 30 apply.
30 - Anesthesia machine is equipped with inhalation anesthesia vaporizers without an agent analyzer
in the breathing system. Code B.
The
correct
certification
RECOMMENDATIONS.
A Narkomed 6000 could have no DMI recommendations or failure codes apply. The correct
certificaiton level for this machine is Code A, CERTIFIED.
Code D, which means NO CERTIFICATION, also means the machine shall not receive a
Periodic Manufacturer's Certification label. The machine shall also receive a 'WARNING -
This System is Not Certified' label, P/N 4114857. This label shall be placed at a prominent
location on the right side of the machine after all other previous PM and 'Vigilance Audit®
Validation' labels have been removed.
Rev.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
for
this
machine
is
for
this
machine
is
6-1H
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)
B,
which
means
B,
which
means
CERTIFIED
WITH
CERTIFIED
WITH

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents