PMC PROCEDURE (continued)
6.21.19
Using a stopwatch, time the inspiratory phase.
( )
6.21.20
What is the inspiratory time? ___ seconds (1.8-2.2)
6.21.21
Using a stopwatch, time the expiratory phase.
( )
6.21.22
What is the expiratory time? ___ seconds (3.6-4.4)
**
6.21.23
Press and hold the EXTENDED RANGE switch and scroll the I:E ratio
dial counter-clockwise and verify the extended I:E values increment
(2:1, 3:1 and 4:1); return the I:E ratio to 2:1.
**
6.21.24
Using a stopwatch, time the inspiratory phase.
( )
**
6.21.25
What is the inspiratory time? ___ seconds (3.6 - 4.4)
**
6.21.26
Using a stopwatch, time the expiratory phase.
( )
**
6.21.27
What is the expiratory time? ___ seconds (1.8 - 2.2)
( )
6.21.28
Adjust the FREQUENCY and I:E RATIO through the following settings
and verify that the ventilator cycles properly:
FREQ.
11
22
33

6.22 BELLOWS DRIVE GAS LEAK: ADULT

6.22.1
Remove the ventilator hose from the VENTILATOR HOSE terminal.
6.22.2
Attach a test terminal to the bellows assembly ventilator hose terminal.
6.22.3
Connect a flowmeter test stand (P/N S000081) to the test terminal.
6.22.4
Set the FREQUENCY to 1 BPM.
6.22.5
Set the I:E RATIO to 1:1.
6.22.6
Set the INSPIRATORY FLOW to the maximum.
( )
6.22.7
What is the flow indicated during the inspiratory phase? ___ (<50 ml)
6.22.8
Remove the test terminal and flowmeter test stand.
6.22.9
Turn the ventilator OFF.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
I:E RATIO
FREQ.
1:1
44
1:1.5
55
1:2
66
6-50
I:E RATIO
FREQ.
1:2.5
77
1:3
88
1:3.5
99
NM2B
I:E RATIO
1:4
1:4.5
1:4.5
Rev. Y