Loss Of Breathing Pressure Monitor - Dräger Narkomed 2B Service Manual

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

NM2B
START
CONNECT
N
PILOT LINE
AT BOTH ENDS
CONNECT TUBING TO
TRANSDUCER ON
PROCESSOR BOARD
N
(ALSO CHECK
CONTINUITY OF
INTERNAL TUBING
CONNECTIONS)
Rev. A
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
GUIDE 2: Loss of Breathing Pressure Monitor
IS
Y
BREATHING
PRESSURE MONITOR
FUNCTIONAL?
N
IS
PILOT LINE
CONNECTED TO
INTERFACE PANEL AND TO
PATIENT Y-PIECE OR
ABSORBER?
Y
IS
BREATHING
PRESSURE TUBING
CONNECTED TO
TRANSDUCER ON
PROCESSOR
BOARD?
Y
REPLACE
PROCESSOR ASSEMBLY
AS OUTLINED IN
PROCEDURE 4.26
3-5
RETURN TO CD-ROM TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued)
IS
Y
BREATHING
PRESSURE MONITOR
FUNCTIONAL?
N
REINSTALL ORIGINAL
PROCESSOR
ASSEMBLY
CONTACT NAD
SERVICE DEPT.
PERFORM A
COMPLETE PMS
ON UNIT
N
DOES
UNIT PASS
PMS?
Y
UNIT IS
FUNCTIONAL