A Oxygen Ratio Controller - Dräger Narkomed 2B Service Manual

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

PMC PROCEDURE (continued)
6.19A
OXYGEN RATIO CONTROLLER - if applicable
*
6.19A.1
Set the Gas Selector switch to O
6.19A.2
Depress the O
6.19A.3
Set the oxygen flow to 1000 ml/min.
6.19A.4
Open the nitrous oxide flow control valve to the stop position.
( )
6.19A.5
What is the oxygen concentration after the value stabilizes? ___ % (21-
29)
6.19A.6
Adjust the oxygen flow to 1.5 l/min.
( )
6.19A.7
What is the oxygen concentration after the value stabilizes? ___ % (21-
29)
6.19A.8
Adjust the oxygen flow to 2 l/min.
( )
6.19A.9
What is the oxygen concentration after the value stabilizes? ___ % (21-
29)
6.19A.10 Adjust the oxygen flow to 4 l/min.
( )
6.19A.11 What is the oxygen concentration after the value stabilizes? ___ % (21-
29)
6.19A.12 Reduce the O
than or equal to 600 ml/min.
6.19A.13 Close the oxygen flow control valve.
( )
6.19A.14 What is the flow of nitrous oxide? ___ ml/min. (0) or 375-750 ml/min on
machines with later design ORC (P/N 4113329-001)
6.19A.15 What is the oxygen concentration with the O
closed?___% (>21)
6.19A.16 Close the nitrous oxide flow control valve.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
+ N
2
FLUSH for 5 seconds.
2
flow to 500 ml/min. Verify that the N
2
6-44
O.
2
O flow is greater
2
flow control valve
2
NM2B
Rev. Y