6.23 SPIROMED

6.23.1
Press the LOW key in the Minute Volume Alarms section of the main
keypad. A box shall appear around the Minute Volume Alarm Limit.
6.23.2
What is the low minute volume alarm default? ___ (3.0)
6.23.3
Verify that the minute volume has a low alarm limit range from 0.5 to
10.0 by increments of 0.1.
6.23.4
Adjust the low minute volume alarm to 2.0 liters.
6.23.5
Disable the VOLUME ALARM by pressing the VOLUME ALARMS
DISABLE key in the Minute Volume Alarms section of the main
keypad. Turn on the ventilator and start a stopwatch.
( )
6.23.6
What is the time when APNEA-VOLUME appears in the Caution
column? ___ sec (13-17)
( )
6.23.7
What is the time when APNEA-VOLUME appears in the Warning
column? ___ sec (26-34)
( )
6.23.8
After one (1) minute, does the MINUTE VOLUME LOW message
appear in the Caution column? ___ Y
6.23.9
Insert a test minute volumeter in between the Spiromed sensor and the
exhalation valve.
6.23.10
Attach a patient circuit with a 3-liter bag to the absorber system.
6.23.11
Reconnect the ventilator hose to the Ventilator Hose terminal.
6.23.12
Adjust the FREQUENCY to 6 BPM.
6.23.13
Adjust the I:E RATIO to 1:2.
6.23.14
Adjust the flow to the maximum of the LOW zone.
6.23.15
Adjust the oxygen flow to 2 l/min.
6.23.16
Adjust the Tidal Volume to 200 ml.
Rev. Y
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)
6-51