Command Descriptions
Syntax
Arguments
Returns
Examples
[SOURce[1]]:SWEep:HTIMe
Group
Syntax
Arguments
Returns
Examples
[SOURce[1]]:SWEep:MODE
2-70
[SOURce]:ROSCillator:SOURce
[SOURce]:ROSCillator:SOURce?
INTernal means that the reference clock is set to internal.
EXTernal means that the reference clock is set to external.
INT|EXT
SOURce:ROSCillator:SOURce INTernal
selects the internal clock reference.
This command sets or queries the sweep hold time. Hold time represents the
amount of time that the frequency must remain stable after reaching the stop
frequency.
Source
[SOURce[1]]:SWEep:HTIMe
[SOURce[1]]:SWEep:HTIMe
<seconds>::=<NRf>[<units>]
where:
<NRf> is the hold time in seconds.
<units>::=[ns | μs | ms | s]
<seconds>
SOURce1:SWEep:HTIMe 1ms
sets the CH 1 hold time to 1 ms.
The [SOURce[1]]:SWEep:MODE command selects auto or manual for the sweep
mode for the specified channel.
AFG2021 Arbitrary Function Generator Programmer Manual
{INTernal|EXTernal}
{<seconds>|MINimum|MAXimum}
?
{MINimum|MAXimum}